Calibrate TXP-WTA Sensor Transmitters that communicate to the TracXP wireless monitoring system with a
known concentration of gas at start-up. Check calibration on a regular schedule (at least every 90 days).
More frequent inspections are encouraged to spot problems such as dirt, oil, paint, grease or other foreign
materials on the sensor filter.
•
Do not paint any part of the TXP-WTA Sensor Transmitter or corresponding components.
•
Do not use the TXP-WTA Sensor Transmitter if any enclosure is damaged or cracked or has missing
components.
•
Make sure covers, internal boards, and antenna connections are securely in place before operation.
•
Use only approved Smart Sensor Modules that are compatible with the TXP-WTA Sensor Transmitter.
•
Periodically test for correct operation of the system’s alarm events by exposing the TXP-WTA Sensor
Transmitter to a known concentration of gas above the High Alarm set-point.
•
Do not expose TracXP devices to electrical shock or continuous severe mechanical shock.
•
Protect TracXP devices from dripping liquids and high-power sprays.
•
Use only for applications described within this manual.