Manual
Instructions for use
BreastBoards & ThoraxSupports
5
03062022V14IFU122500
MacroMedics ®
2.2
MR CONDITIONAL
The following devices are MR Conditional:
• 122150
ThoraxSupport™, MRC
• 122310
Upper AS TS/MBLX, MRC, set/2
• 122340 Cranial AS for BBLX/TS MRC, set/2
• 122660
ThoraxSupport™, short MRC
• 122710
Rotatable AS LV MRC, set/2
• 122740 Armcup for handpoles, set/2
• 122750 BreastBoard LX™, MRC
• 126190
OmniBoard™ bottomstop MRC
• 174590
OmniBoard™ BTM-SX-RE, MRC
• 175060
OmniBoard™ BTM-SX-S, MRC
The devices labelled as MR Conditional can be safely used in a MR system meeting the following
conditions:
•
Static magnetic fi eld of 1.5 Tesla and 3 Tesla,
with
•
Maximum spatial fi eld gradient of 2,500 G/cm(25 T/m)
• Maximum force product of 50,000,000 G²/cm (50 T
2
/m)
Due to size and location of the metallic parts, no clinically signifi cant RF heating is expected.
It is the responsibility of the user to ensure that these conditions are met.
2.3
MR SAFE
The following devices are MR Safe:
• 122530
WedgeThoraxSupport™, MR
3. Intended Use and Indications for Use
Intended Use
: Positioning of the patient during radiotherapy and radiodiagnostics, including MR where
indicated.
Indications for Use
: Patients aged 18 or older undergoing radiation therapy of the breast and chest
region, including photon, proton and electron treatments. This includes positioning of the patient during
image acquisition to support treatment.