NOTICE
Hazard due to laser radiation: The barcode scanner of the
NANOCOLOR
®
spectrophotometer is subject to laser protection class 1M. Do not look directly into
the beam of light from the barcode reader.
LASER RADIATION
DO NOT LOOK AT DIRECTLY
WITH OPTICAL INSTRUMENTS
LASER CLASS 1M
Rear view
Fig. 1: Rear view
UV
/
VIS
II
The
NANOCOLOR
®
UV
/
VIS
II requires voltage of
110 / 230 V (~ 50 / 60 Hz). Insert the power plug into
the power connection (Fig. 1
) on the back of the
device.
Fig. 2: Rear view
VIS
II
The
NANOCOLOR
®
VIS
II requires voltage of 12V
(min. 3.5A). Insert the corresponding power adapter
on the mains adapter and then connect the power
supply unit to the power connection (Fig. 2
back of the device.
Insert the power plug into an outlet.
A
USB A connection
B
USB B connection
C
Ethernet (LAN) interface
D
RS 232 interface
E
Power connection 110 / 230 V
(~ 50 / 60 Hz)
F
On/Off switch
G
SDHC card slot (
UV
/
VIS
II underneath the lamp
cover; see the operating instructions Fig. 1d)
H
Power connection 12 V DC 3.5A
NOTICE
For the
VIS
II spectrophotometer, only the associated mains adapter GTM96600‑6012‑R2
(REF 919156) may be used.
The on/off switch of the
NANOCOLOR
®
UV
/
VIS
) /
NANOCOLOR
®
VIS
) is
on the back of the device. Switch the device on. After it is switched on, a start screen showing
the manufacturer’s logo appears. The device performs a function test lasting approx. 1 minute.