6.1.5 Transmission
The basic function “Transmission” enables the measurement of a sample’s
transmission against a zero solution. After having selected this method, a win-
dow will appear asking for the wavelength of measurement. By pressing the
Add
button a set of wavelengths can be added to the list. It is possible to add up
to 10 wavelengths to the list. The wavelenghts list can be cleared or selected entries can
be deleted by pressing the
Remove
button. Confirming with
will open the mea-
surement window. A request for a zero solution will appear. Inserting the zero solution
and pressing
initiates the zero measurement. After the successful measurement of
the zero solution, the sample solution needs to be inserted. Pressing
starts the
measurement process. After the measurement has been completed, the measurement
result will be displayed in the measurement window.
It is possible to perform more than one measurement in a row. The results are then listed
in a list on the left side of the result window.
Detailed information for each measurement can be called up by choosing an entry from
the list.
6.1.6 Turbidity
The basic function “Turbidity” enables the measurement of a samples nephelo-
metric turbidity. Insertion of the test tube and pressing
initiates the measure-
ment. After the measurement has been completed, the measurement result will
be displayed in the result window.