Instruction Manual
MAN-001
18
www.macmedical.com
Operation of Display Panel
Start
Press the power switch to the “ON” position. (For multi-
chamber units, each chamber has its own power
switch). The upper readout display will show the
actual chamber temperature.
Set Processing Temperature
1. Press “SET” and release.
2.
Press the UP ▲ arrow key up (to raise the setpoint)
or the down ▼ arrow key (to lower the setpoint).
The setpoint is indicated in the lower row of the
display.
3. Press “SET” again to complete the change (If dual
or triple chamber unit, steps must be completed
for all chambers). The setpoint temperature will
change and the last digit will flash.
4.
To prevent adjustment to the setpoint
temperature, the control may be locked with the
Key Switch (see #9, Fig. 10)
NOTE:
See the Maximum/Minimum Limits Temperature
Set Points table below for various warming cabinets
and chambers.
NOTE: The temperature setting may be changed
at any time. However, if the setpoint is changed
more than 10° below the actual temperature, the HI
temperature alarm will activate.
If alarm is activated and the control panel reads “HI”,
complete the following steps:
1.
Press “ALARM RESET” (UP ▲ arrow) until the alarm
stops.
2. Open door(s) to release heated air.
3. If the display reads “HI” and alarm remains, open
door(s) and allow more cooling time.
NOTE:
If the alarm is activated under normal
operating conditions, turn power to specific chamber
off and call your Mac Medical representative at
1-877-828-9975.
Caution
Burn Hazard:
DO NOT raise the
setpoint temperature to increase the rate of heating.
This could overheat the contents leading to possible
patient burns.
Warning
Explosion Hazard:
DO NOT exceed
150° for non-vented closures (screw caps, crimp seals,
plastic pouches, etc. DO NOT exceed pre-sterile
solution manufacturer’s temperature requirements.
Minimum/Maximum limits Temperature Set Points and Temperature Tolerances on Warming Cabinet Units
Warmer Cabinet Model
Minimum
Temperature
Set Point
Maximum
Temperature
Set Point
blankets
Irrigation
fluids
Injection
fluids
Temperature
Tolerance
SWC151822
90°F (32°C)
135°F (57°C)
30°C to 71°C
(86°F to 160°F)
30°C to 66°C
(86°F to 150°F)
30°C to 40°C
(86°F to 104°F)
±1°F
Single chamber units SWC182424, SWC182436, SWC183024,
SWC182464, SWC242436, SWC243024, SWC242424
160°F (71°C)
±1°F
Single chamber units SWC183036, SWC243036
±2°F
Single chamber units SWC182474, SWC183064, SWC183074,
SWC242464, SWC242474, SWC243064, SWC243074
±3°F
Dual Chamber units (Upper Chamber) DWC182464T, DWC183064T,
DWC182474T, DWC183074T, DWC242474T, DWC243074T
±1°F
Dual Chamber units (Upper Chamber) DWC182474E, DWC183064T,
DWC183074E, DWC242464T, DWC242474E, DWC243064T, DWC243074E
±2°F
Dual Chamber units (Lower Chamber) DWC182464T, DWC182474E,
DWC183064T, DWC182474T, DWC183074T, DWC183074E,
DWC242464T, DWC242474E, DWC243064T, DWC243074E
±2°F
Dual Chamber units (Lower Chamber) DWC242474T, DWC243074T
±3°F
Triple chamber models
110°F (43°C) top
chamber
±1°F
135°F (57°C)
middle chamber
±1°F
160°F (71°C)
bottom chamber
±1°F