R1M-GH

CHECKING

1)  Terminal wiring:  Check that all cables are correctly con-

nected according to the connection diagram.

2)  Power input: Check supply voltage.
3)  Input:  Check that the input is within 0 – 100% of full-

scale.

4)  DIP SW setting: Check that all switches are set correct-

ly.

ADJUSTMENT PROCEDURE

This  unit  is  calibrated  at  the  factory  to  meet  the  ordered 
specifications,  therefore  you  usually  do  not  need  any  cali-
bration.

INPUT RANGE SELECTING

Select switches to match the input signal according to the table below.

M-SYSTEM WARRANTY

M-System 

 

 
 
 
 

 
 

SW311 SW305 SW304 SW303 SW302

SW405 SW404 SW403 SW402

SW301

O
N

2

1

3
4
5
6
7
8

SW702

SW702 is fixed to 
the position 3.
DO NOT CHANGE.

SW411

SW401 SW308 SW307 SW306

SW408 SW407 SW406

O
N

2

1

3
4
5
6
7
8

MAINTENANCE

Regular calibration procedure is explained below:

 CALIBRATION

Warm up the unit for at least 10 minutes.  Apply 0%, 25%, 
50%,  75%  and  100%  input  signal.    Check  that  the  output 
signal for the respective input signal remains within accu-
racy described in the data sheet.  When the output is out of 
tolerance, please contact M-System’s Sales Office or repre-
sentatives.

INPUT

SWITCHES

SETTING

NOTES

Thermocouple

SW301 – SW308  (ch 1 – ch 8)

SW401 – SW408  (ch 9 – ch 16)

Set to position “1”

                                

Attenuation OFF

SW311  (ch 1 – ch 8)

SW411  (ch 9 – ch 16)

OFF

Voltage input, 0.8V or less

SW301 – SW308  (ch 1 – ch 8)

SW401 – SW408  (ch 9 – ch 16)

Set to position “1”

                                

Attenuation OFF

SW311  (ch 1 – ch 8)

SW411  (ch 9 – ch 16)

ON

Voltage input, above 0.8V (*)

SW301 – SW308  (ch 1 – ch 8)

SW401 – SW408  (ch 9 – ch 16)

Set to position “3”

                                

Attenuation ON

SW311  (ch 1 – ch 8)

SW411  (ch 9 – ch 16)

ON

(*) Factory default setting

Rugghölzli 2
CH - 5453 Busslingen

Tel. +41 (0)56 222 38 18

Fax +41 (0)56 222 10 12

[email protected]

www.sentronic.com

Produkte, Support und Service

SENTRONIC

AG

Summary of Contents for R1M Series

Page 1: ...rotocol Modbus RTU Protocol Modbus RTU Refer to Modbus Protocol Reference Guide EM 5650 for supported functions RS 232 C Standard Conforms to RS 232 C EIA Transmission distance 10 meters max RS 485 St...

Page 2: ...700 C or 1292 F The described accuracy may be partially not satisfied when the temperature ranges below 0 C STANDARDS APPROVALS CE conformity EMC Directive 2004 108 EC EMI EN 61000 6 4 2007 EMS EN 61...

Page 3: ...e When the device is not at the end remove the jumper pin Remark 1 This device is not designed to cancel noise included in the input signals Be careful to eliminate such noise by using shielded cables...

Page 4: ...slave 04 Read Input Registers X Collected data from the field by the slave 05 Force Single Coil X Digital output from the slave 06 Preset Single Registers X General purpose register within the slave...

Page 5: ...17 48 F Reserved for AO in engineering unit 145 160 Y Y I I O type No 161 176 Y I Burnout type 161 168 for Type J3 0 No burnout 1 Upscale 2 Downscale 4 I 16 bit integer F 32 bit floating B16 16 byte...

Page 6: ...E A D DATA Decimal DC 20V 20000 5V 5000 1V 10000 0 8V 8000 0 2V 20000 50mV 5000 10mV 10000 Thermocouple Temperature x 10 RTD Temperature x 10 Potentiometer 0 100 2000 0 500 0 1k 0 10k BIT NAME DESCRIP...

Page 7: ...00 to 600 C 0X1B L 200 to 900 C 0X1C P Platinel II 0 to 1395 C R1M J3 RTD input 0X30 JPt 100 JIS 89 200 to 500 C 0X31 Pt 100 JIS 89 200 to 660 C 0X32 Pt 100 JIS 97 IEC 200 to 850 C 0X33 Pt 50 JIS 81 2...

Page 8: ...istance between the PC and the R1Ms is long insert an RS 232 C RS 485 Converter for isolation RS 232 C RS 232 C RS 485 Converter model R2K 1 Specifications are subject to change without notice Ruggh l...

Page 9: ...urce connected to the unit and TDK Model ZCAT3035 1330 or equivalent for the RS 232C cable are recommended The equipment must be installed such that appropriate clearance and creepage distances are ma...

Page 10: ...5 T1 T2 T3 T4 T5 T6 T7 CJC Sensor model CJM RS 485 Connector Address Setting Rotary SW Configurator Jack Specifications Marking Power Terminals RS 232C 9 pin Connector RUN Indicator LED DIN Rail Adapt...

Page 11: ...t the end remove the jumper pin Remark 1 This device is not designed to cancel noise included in the input signals Be careful to eliminate such noise by using shielded cables Remark 2 Be sure to maint...

Page 12: ...alibration procedure is explained below CALIBRATION Warm up the unit for at least 10 minutes Apply 0 25 50 75 and 100 input signal Check that the output signal for the respective input signal remains...

Reviews: