background image

R1M-J3

P.  / 4

EM-5654 Rev.5

PC RECORDER

(RTD or potentiometer input, 8 points)

MODEL

R1M-J3

INSTRUCTION MANUAL

BEFORE USE ....

Thank you for choosing M-System.  Before use, please check 
contents of the package you received as outlined below.
If  you  have  any  problems  or  questions  with  the  product, 
please contact M-System’s Sales Office or representatives.

This product is for use in general industrial environments, 
therefore may not be suitable for applications which require 
higher level of safety (e.g. safety or accident prevention sys-
tems) or of reliability (e.g. vehicle control or combustion con-
trol systems).

For  safety,  installation  and  maintenance  of  this  product 
must be conducted by qualified personnel.

 PACKAGE INCLUDES:

  PC Recorder .................................................................. (1)
  Cable (9-pin D-sub cable, straight type) ...................... (1)
  CD (software and users manual) ................................. (1)

 MODEL NO.

Confirm  Model  No.  marking  on  the  product  to  be  exactly 
what you ordered.

 INSTRUCTION MANUAL

This  manual  describes  necessary  points  of  caution  when 
you use this product, including installation, connection and 
basic maintenance procedures.  Read also the Users Manual 
for the software included in the CD for maximum use of the 
PC Recorder.

POINTS OF CAUTION

 CONFORMITY WITH EC DIRECTIVES

• This equipment is suitable for use in a Pollution Degree 

2 environment and in Installation Category II, with the 
maximum operating voltage 300V.

  Basic insulation is maintained between the signal input 

and RS-232C/RS-485.  Prior to installation, check that the 
insulation class of this unit satisfies the system require-
ments.

• Altitude up to 2000 meters
• The equipment must be mounted inside a panel.
• Insert noise filters.  Okaya Electric Industries Model SUP-

E1H or equivalent for the power source connected to the 
unit,  and TDK  Model  ZCAT3035-1330  or  equivalent  for 
the RS-232C cable are recommended.

• The  equipment  must  be  installed  such  that  appropriate 

clearance and creepage distances are maintained to con-
form to CE requirements.

• The actual installation environments such as panel con-

figurations,  connected  devices,  connected  wires,  may  af-
fect the protection level of this unit when it is integrated 
in a panel system.  The user may have to review the CE 
requirements in regard to the whole system and employ 
additional protective measures to ensure the CE conform-
ity.

 POWER INPUT RATING & OPERATIONAL RANGE

Locate the power input rating marked on the product and 
confirm its operational range as indicated below:
  100 – 240V AC rating: 85 – 264V, 47 – 66 Hz, approx. 10VA
  24V DC rating: 24V ±10%, approx. 7W
 

 GENERAL PRECAUTIONS

Before  you  remove  the  module,  turn  off  the  power  supply 
and input signal for safety.

 PC RECORDER SOFTWARE

Use the latest version of PC Recorder Software included in 
the product package.

 ENVIRONMENT

• Indoor use
• When heavy dust or metal particles are present in the air, 

install  the  module  inside  proper  housing  with  sufficient 
ventilation.

• Do not install the module where it is subjected to continu-

ous vibration.  Do not apply physical impact to the mod-
ule.

• Environmental temperature must be within -5 to +60°C     

(23 to 140°F) with relative humidity within 30 to 90% RH 
in order to ensure adequate life span and operation.

• Be sure that the ventilation slits are not covered with ca-

bles, etc.

 WIRING

• Wrong connection may damage the module.
• Do not connect cables to moving parts or pull them tight-

ly.

• Do  not  install  cables  (power  supply,  input  and  output) 

close  to  noise  sources  (relay  drive  cable,  high  frequency 
line, etc.).

• Do  not  bind  these  cables  together  with  those  in  which 

noises are present.  Do not install them in the same duct.

 AND ....

The module is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for sat-
isfying complete performance described in the data sheet.

Summary of Contents for R1M-J3

Page 1: ... for the RS 232C cable are recommended The equipment must be installed such that appropriate clearance and creepage distances are maintained to con form to CE requirements The actual installation environments such as panel con figurations connected devices connected wires may af fect the protection level of this unit when it is integrated in a panel system The user may have to review the CE requir...

Page 2: ...SCSI devices may impair appropriate performance MSR128LS MSR128LV MSR128LS MSR128LV PC IBM PC AT or compatible Operating system Microsoft Windows XP SP1 or SP2 For High Speed Mode Group 0 50 msec storing cycle use Windows XP SP1 SP2 Note Proper software functions may not be ensured under certain conditions CPU Pentium II 233 MHz or higher 2 Screen area 800 by 600 pixels or better resolution 640 by...

Page 3: ...ions therefore you usually do not need any cali bration MAINTENANCE Regular calibration procedure is explained below CALIBRATION Warm up the unit for at least 10 minutes Apply 0 25 50 75 and 100 input signal Check that the output signal for the respective input signal remains within accu racy described in the data sheet When the output is out of tolerance please contact M System s Sales Office or ...

Page 4: ...l terminating resistor is used when the device is at the end of a transmission line 2 Install shielded cables to all sections and ground them at single point 2 FG RS 485 TERMINAL CONNECTIONS Connect the module referring to the connection diagram RS 232C T3 T4 T2 T1 T7 T6 T5 Term Resist Output Circuit When the device is located at the end of a transmission line via twisted pair cable when there is ...

Reviews: