CALIBRATION

This unit is calibrated at the factory to meet the ordered 
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case 
of regular calibration, adjust the output as explained in the 
following.

 

ZERO/SPAN & UP/DOWN SWITCHES

Fine zero and span adjustments can be done with the ZERO/
SPAN Selector and UP/DOWN switches.  Calibrated values 
are stored in the non-volatile memory, which will not be lost 
even when you turn off power supply to the unit.
The output 1 and output 2 can be independently adjustable.  
Zero and span are respectively set to 0% and 100% at the 
factory, and field calibration can be reset to ex-factory set-
ting if necessary.

Adjustable range: 

±5%

Zero/Span Selector

 

ZERO

:  UP/DOWN switches usable for zero adjustment.

 

OFF

:  

UP/DOWN switches unavailable.

 

SPAN

:  UP/DOWN switches usable for span adjustment.

UP/DOWN Switches

 

UP

:   

Pressing UP increases adjusted values.

 

DOWN

:  Pressing DOWN decreases adjusted values.

 

FRONT VIEW

UP Switch

Status Indicator LED

Configurator Jack

DOWN Switch

 

BOTTOM VIEW

Zero/Span Selector

S

Z

OFF

Z

S

Zero/Span Selector

MODE

SW POSITION

SPAN

S

OFF

ZERO

Z

 

HOW TO CALIBRATE THE ZERO & SPAN

Use a signal source and measuring instruments of sufficient 
accuracy level.  Turn the power supply on and warm up for 
more than 10 minutes.
The output 1 and output 2 adjustments are linked except 
for (2) and (6).

1) Apply 0% input.
2) Select the Output 1 Zero: Slide the Zero/Span Selector to 

the position Z.

  Select the Output 2 Zero: Slide the Zero/Span Selector to 

the position Z while pressing UP switch.

3) Press UP or DOWN switch and adjust the output to 0%.  

Incrementing speed will be doubled when you keep press-

ing a switch.

4) Put the Zero/Span Selector back to the position OFF.
5) Apply 100% input.
6) Select the Output 1 Span: Slide the Zero/Span Selector to 

the position S.

  Select the Output 2 Span: Slide the Zero/Span Selector to 

the position S while pressing UP switch.

7) Press UP or DOWN switch and adjust the output to 

100%.  Incrementing speed will be doubled when you 

keep pressing a switch.

8) Put the Zero/Span Selector back to the position OFF.
9) Check Zero and Span adjustments again with 0% and 

100% input respectively.

10) If the output value is changed, repeat the above proce-

dure (1) – (9).

 

HOW TO RESET

After you calibrated manually with these switches, you 
can reset them to its ex-factory state by pressing both UP/
DOWN switches at once.

Summary of Contents for M8XV2

Page 1: ...firm its operational range as indicated below 24V DC rating 24V 10 approx 50mA 70mA for current output GENERAL PRECAUTIONS Before you remove the unit or mount it turn off the power supply and input signal for safety ENVIRONMENT Indoor use When heavy dust or metal particles are present in the air install the unit inside proper housing with sufficient ventilation Do not install the unit where it is ...

Page 2: ...ase OUTPUT 1 6 7 8 9 10 11 OUTPUT 2 1 2 A B POWER INPUT CONFIGURATOR JACK CHECKING 1 Terminal wiring Check that all cables are correctly con nected according to the connection diagram 2 DIP switches Check the input range selector and output range selector setting 3 Power input Check the power input voltage 4 Input Check that the input signal is within 0 100 of the full scale 5 Output Check that th...

Page 3: ...figurator Jack DOWN Switch BOTTOM VIEW Zero Span Selector S Z OFF Z S Zero Span Selector MODE SW POSITION SPAN S OFF ZERO Z HOW TO CALIBRATE THE ZERO SPAN Use a signal source and measuring instruments of sufficient accuracy level Turn the power supply on and warm up for more than 10 minutes The output 1 and output 2 adjustments are linked except for 2 and 6 1 Apply 0 input 2 Select the Output 1 Ze...

Page 4: ...STATUS INDICATOR LED The unit is provided with a status indicator LED which blinks in different patterns indicating various status of its CPU When it is functioning within normal parameters the LED blinks in a regular pattern of ON and OFF The following figure indicates typical patterns LED PATTERNS STATUS Normal Programming Mode Out 1 Calibration Mode Out 2 Calibration Mode 200 ms ON OFF MAINTENA...

Reviews: