background image

M3SMS

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201  Fax: +81(6)6659-8510  E-mail: [email protected]

EM-7715  Rev.4  P. 3 / 4

TERMINAL CONNECTIONS

Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.

 

EXTERNAL DIMENSIONS  unit: mm (inch)

7

8

5

6

1

2

3

4

106 (4.17)

12

(.47)

110.5 (4.35)

62 (2.44)

[5 (.20)]

• When mounting, no extra space is needed between units.

DIN RAIL

35mm wide

 

CONNECTION DIAGRAM

+

1

4

3

2

1

POTENTIOMETER

max.

min.

3

+

OUTPUT

5

6

U(+)

V(–)

POWER

7

8

 

WIRING INSTRUCTIONS

  • Applicable wire size

   Solid:  

0.2 to 2.5 mm

2

 (0.55 to 1.75 dia.)

   Stranded:  0.2 to 2.5 mm

2

      

 

 Tinning wire ends may cause contact failure  

and therefore is not recommended.

     Ferruled:  0.2 to 1.5 mm

2

 (0.55 to 1.35 dia.)

      

 

 The following Phoenix Contact terminals are  

recommended:

      

 

AI 0,25-8YE 

0.2 to 0.25 mm

2

      

 

AI 0,34-8TQ 

0.25 to 0.34 mm

2

      

 

AI 0,5-8WH 

0.34 to 0.5 mm

2

      

 

AI 0,75-8GY 

0.5 to 0.75 mm

2

      

 

AI 1,0-8RD 

0.75 to 1.0 mm

2

      

 

AI 1,5-8BK 

1.0 to 1.5 mm

2

  • Expose wire conductors by 8 mm (0.31”).

8 mm

8 mm

4 mm dia.

max.

4 mm dia.

max.

Wire exposure

Recommended
ferruled wire

CHECKING

1) Terminal wiring:  Check that all cables are correctly con-

nected according to the connection diagram.

2) Power input voltage:  Check voltage across the terminal 

7 – 8 with a multimeter.

3) Input:  Check voltage across the terminal 4(+) – 3(–) with 

a voltmeter to show 0V at 0% potentiometer input and 

the same voltage as that across 1(+) – 3(–) at 100% input.

4) Output:  Check that the load resistance meets the de-

scribed specifications.

ADJUSTMENT PROCEDURE

This unit is calibrated at the factory with the total resist-
ance input, therefore you do not need any calibration if you 
use the potentiometer’s total resistance.
When you don’t use the total resistance or in case of regular 
calibration, adjust the output as explained in the following.

 

HOW TO CALIBRATE THE OUTPUT SIGNAL

Use a signal source and measuring instruments of sufficient 
accuracy level.  Turn the power supply on and warm up for 
more than 10 minutes.

1) ZERO:  Apply 0% input and adjust output to 0%.
2) SPAN:  Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-

dure 1) – 3).

Reviews: