42
M3LU2 OPERATING MANUAL EM-2653-B Rev.0
12.1.6 sTANDARDs & APPROVALs
EU conformity
EMC Directive (2004/108/EC)
EMI EN 61000-6-4: 2007/A1: 2011
EMS EN 61000-6-2: 2005
Low Voltage Directive (2006/95/EC)
EN 61010-1: 2010
Installation Category II
Pollution Degree 2
Input or output to power: Reinforced insulation (300 V)
Input to output: Basic insulation (300 V)
12.1.7 I/O TYPE, MAXIMUM RANGE & ACCURACY
[Table 1] DC input
INPUT TYPE
MINIMUM SPAN
MAXIMUM RANGE
ACCURACY
*
1
DC Current
1 mA
0 to 20 mA DC
±0.1%
DC Millivolt
4 mV
-1000 to +1000 mV DC
±10 μV at F.S. input ≤ 50 mV
±40 μV at F.S. input ≤ 200 mV
±60 μV at F.S. input ≤ 500 mV
±80 μV at F.S. input > 500 mV
DC Voltage
1 V
-10 to +10 V DC
±0.1%
*
1 Or ±0.1% of span, whichever is greater.
[Table 2] Thermocouple input
T/C
°C
°F
MIN.
SPAN
MAXIMUM
RANGE
ACCURACY
*
2
CONFORMANCE
RANGE
MIN.
SPAN
MAXIMUM
RANGE
ACCURACY
*
2
CONFORMANCE
RANGE
(PR)
20
0 to 1760
±1.00
0 to 1760
36
32 to 3200
±1.80
32 to 3200
K (CA)
20
-270 to +1370
±0.25
-150 to +1370
36
-454 to +2498
±0.45
-238 to +2498
E (CRC)
20
-270 to +1000
±0.20
-170 to +1000
36
-454 to +1832
±0.36
-274 to +1832
J (IC)
20
-210 to +1200
±0.25
-180 to +1200
36
-346 to +2192
±0.45
-292 to +2192
T (CC)
20
-270 to +400
±0.25
-170 to +400
36
-454 to +752
±0.45
-274 to +752
B (RH)
20
100 to 1820
±0.75
400 to 1760
36
212 to 3308
±1.35
752 to 3200
R
20
-50 to +1760
±0.50
200 to 1760
36
-58 to +3200
±0.90
392 to 3200
S
20
-50 to +1760
±0.50
0 to 1760
36
-58 to +3200
±0.90
32 to 3200
C (WRe 5-26)
20
0 to 2315
±0.25
0 to 2315
36
32 to 4199
±0.45
32 to 4199
N
20
-270 to +1300
±0.30
-130 to +1300
36
-454 to +2372
±0.54
-202 to +2372
U
20
-200 to +600
±0.20
-200 to +600
36
-328 to +1112
±0.36
-328 to +1112
L
20
-200 to +900
±0.25
-200 to +900
36
-328 to +1652
±0.45
-328 to +1652
P (Platinel II)
20
0 to 1395
±0.25
0 to 1395
36
32 to 2543
±0.45
32 to 2543
*
2 [Ac Cold Junction Compensation Error 1.0°C (1.8°F)] or ±0.1% of span, whichever is greater.