background image

M3LPA

P.  / 9

EM-2656 Rev.4

 I/O RANGE LABEL

Blank I/O range labels are included in the product pack-
age.  Write in the configured ranges and put the label on 
the side as shown below.

INPUT

OUTPUT

I/O Range Label (included in the product package)

 OUTPUT TYPE & RANGE 

Table 14

OUTPUT TYPE 

MINIMUM SPAN 

MAXIMUM RANGE 

CONFORMANCE RANGE

DC Current 

1mA 

0 to 20mA 

0 to 24mA 

 

DC Voltage, Narrow Spans 

250mV 

-2.5 to +2.5V 

-3 to +3V 

 

DC Voltage, Wide Spans 

1V 

-10 to +10V 

-11.5 to +11.5V

 INPUT TYPE & RANGE 

Table 13

INPUT TYPE 

MINIMUM AMPLITUDE 

MAXIMUM RANGE 

*

4

 

MAXIMUM AMPLITUDE

Open Collector 

4V 

0 – 100 kHz 

12V 

Mechanical Contact 

4V 

0 – 10 Hz 

12V 

 

Voltage Pulse 

0.1V 

0 – 100 kHz 

100V

Two-wire Current Pulse 

4mA 

0 – 100 kHz 

25mA 

 

RS-422 Line Driver Pulse 

–– 

0 – 100 kHz 

––

*4. Minimum span: 10% of the maximum frequency range selected in Table 4.  (Max./Min. pulse width: 10 sec. / 10 µsec.)

 

Time constant is set to 50 msec. with the ‘large’ noise filter setting and to 10 msec. with the ‘small’ noise filter setting 

(Table 10).

 ZERO & SPAN ADJUSTMENTS

After the transmitter is installed and operational, fine zero 
and span tuning can be performed as explained below.  Both 
zero and span are adjustable within ±15%.
1)  Run  Mode:  Confirm  that  the  LD1  green  light  is  blink-

ing.

2)  Fine Zero Calibration Mode: Press MODE button for 1 

or 2 seconds until the LD1 red light is ON and the LD2 
green light is blinking.

3)  Use UP (increase) and DOWN (decrease) buttons to ad-

just the output to 0%.

4)  Fine  Span  Calibration  Mode:  Push  MODE  button  and 

confirm that the LD1 red light is ON and the LD3 green 
light instead of LD2 is blinking.

5)  Use UP (increase) and DOWN (decrease) buttons to ad-

just the output to 100%.

6)  Run Mode: Push MODE button and confirm that only the 

LD1 green light is blinking.

Note 1: Calibration steps can be skipped when not needed 
by repeating pushing MODE buttons.
Note 2: There is no limitation of entering values for multi-
ple times within one step of Calibration Mode.  Signal level 
is stored each time the respective UP or DOWN button is 
pressed.  However, for the input ranging, end always with 
100% (UP).
Note 3: Once the transmitter enters the fine zero or span 
calibration mode, the output goes to preadjusted 0% or 100% 
value respectively regardless of the present input signal.

Summary of Contents for M3LPA

Page 1: ...e filters for the power source connected to the unit Densei Lambda Model MBS 1220 22 or equivalent is recommended The actual installation environments such as panel con figurations connected devices c...

Page 2: ...ng body pull up the driver and pull out the terminal block INSTALLATION DIN RAIL MOUNTING Set the unit so that its DIN rail adaptor is at the bottom Position the upper hook at the rear side of the uni...

Page 3: ...power and DC power connect to different terminals 2 DC POWER INPUT 1 U V 1 3 2 6 5 4 CONFIGURATOR JACK 7 8 10 12 OUTPUT AC POWERED TYPE 2 AC POWER WIRING INSTRUCTIONS Applicable wire size Solid 0 2 to...

Page 4: ...e selected with the DIP SW4 1 through SW4 3 See Tables 9 and 10 For detailed information on the PC configuration refer to the M3CON instruction manual EXTERNAL INTERNAL VIEWS CONFIGURATION MODE DIP SW...

Page 5: ...LD1 green light is blink ing 2 Threshold Calibration Mode Press MODE button for longer than 5 seconds until the LD1 red light is ON 3 Threshold Setting Apply the full scale input level e g 100 Hz from...

Page 6: ...WN UP MODE PUSH DOWN PUSH UP When you set 0 or 100 input output ranges keep pressing UP or DOWN button until the LD1 flashes for approx 5 seconds and turns off which indicates the setup is complete Wh...

Page 7: ...itter is installed and operational fine zero and span tuning can be performed as explained below Both zero and span are adjustable within 15 1 Run Mode Confirm that the LD1 green light is blink ing 2...

Page 8: ...nto voltage V e g Offset 10mA 10mA x 1 10 1V Table 14 PULSE AMPLITUDE SENSITIVITY SCALE 50 100V p p 1 20 25 50V p p 1 10 10 25V p p 1 5 5 10V p p 1 2 1 5V p p 1 0 5 1V p p 1 0 1 0 5V p p 1 A specific...

Page 9: ...g the date that such product was originally purchased if such product has been used under normal operating conditions and properly maintained M System s sole liability and purchaser s exclusive remedi...

Reviews: