![background image](http://html1.mh-extra.com/html/m-system/libo-8/libo-8_instruction-manual_3096845005.webp)
5
Technical data:
-
Supply Voltage: /- 10%, 50Hz
-
Degree of Protection: IP2x (DIN 40050)
-
Working- and Storage Temperature Range: -20 to +80 degree C
-
Housing size : 120mm (L), 65mm (W), 55mm (H)
-
German mains plug and plug socket, other countries upon request
-
Arc Detection starts from about 40W of arc power
-
Cable length of the toroid sensor cable: about 25 cm
-
Direct current of the cable to be monitored:
Toroid coil -8 : minimum of 50A
Toroid coil -16: minimum of 100A (higher currents upon request)
-
Plug socket output for loads, mains switched to plug socket by DPST relay
-
Protection Earth (PE) of plug connected to PE of plug socket
-
Factory-made fuse: 5x20mm, 0,4A, slow blow, inside the housing
Attention: please read carefully chapter 8 for safety advices!
-
Galvanic isolation between cable (Device under Test, DUT) and measuring unit
-
Functionality as per E DIN EN 63027
-
EG-Declaration of conformity available