LC / LCS

P. 2 / 4

EM-1355 Rev.3

DISPLAY SETTING PROCEDURE

  Option /E2

Normal Operating

Mode

Display Configuration

Mode

ITEM : ‘CFG’

Select engineering

unit

ITEM : ‘UNIT’

NEXT

NEXT

NEXT

ENTER

Press NEXT one 

or more times.

ENTER

Select decimal point 

position

ITEM : ‘DOT P’

NEXT

Press NEXT one 

or more times.

ENTER

Calibration Mode

Adjust the display 0% & 

100% values

ITEM : ‘CAL’

Set the present input value as 0%.

Display shows AD conversion value

*

1

.

ITEM : ‘ADJ Z’

NEXT

ENTER

Press NEXT to read and show 

the present input value

(AD conversion value)

Press NEXT to read and show 

the present input value

(AD conversion value)

ENTER

Set the present input value as 100%.

Display shows AD conversion value

*

1

.

ITEM : ‘ADJ S’

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

Select 0%

scaling value

ITEM : ‘SCL Z’

NEXT

Numeric Input

Mode

Exit

ENTER

Select 100%

scaling value

ITEM : ‘SCL S’

Numeric Input

Mode

Exit

ENTER

NEXT

NEXT

NEXT

Display Initialization

Mode

ITEM : ‘INIT’

ITEM : ‘NO’

Exit Numeric

Input Mode

NEXT

ENTER

Press ENTER to reset 

the display 

configuration to the 

initial state.

ITEM : ‘YES’

Press NEXT to 
choose among

‘0’ –> ‘1’ –> ‘-0’ –> 

‘-1’ –> ‘0’ ....

Most Significant Digit

(upper line)

Press NEXT to 

choose among ‘0’ 

through ‘9.’

Press ENTER to 
move to the next 

LS digit.

Less Significant Digits

(upper line)

NEXT

NEXT

The range is set to
10000 or -10000 
and the numeric 
input mode ends.

Select ‘1’ or ‘-1’

Select
‘0’ or ‘-0’

Numeric Input Mode

Press ENTER key in maintaining NEXT key 
pressed to return to Normal Operating Mode 
from any setting mode. 

Select among %, 

µ

V, mV, V, mA, 

°

C, 

°

F, 

, DEG K, mHz, Hz, kHz, 

VAC, AAC, mg, g, kg, t, rpm and rps.

Apply the input signal for 0%.

Apply the input signal for 100%.

(Value Set)

(Value Set)

‘AD conversion value’ means the internal digital data representing 
the input signal, which is computed to a percentage.
It is not a % or engineering unit value.

*

1.

Value

Unit, Item

ENTER Key

NEXT Key

Backlight : Green (normal)

Red (simulated output)