background image

TERMINAL CONNECTIONS

Connect the unit as in the diagram below or refer to the connection diagram on the top of the unit.
Attach the CJC sensor together with input wiring to the input terminals.  The CJC sensor is not interchangeable.  Check that 
its serial number is identical to that of the unit.

 

EXTERNAL DIMENSIONS  unit: mm (inch)

76 (2.99) dia.

40 (1.57)

52.5 (2.07)

1 (.04)

60.9 (2.40)

46 (1.81)

50 (1.97)

4–M3 (.12) TAP

6 (.24) deep

Approx. 10 (.39)

CJC SENSOR (model: CJM)

2 (.08)

1 2 3

5 6

4

r b

 

CONNECTION DIAGRAM

+

T / C

CJC

SENSOR

metal leg

model: CJM

comp. leadwire

1

3

2

+

4

red

black

+

LOAD

6

5

OUTPUT

24V DC

4 – 20mA DC

WIRING INSTRUCTIONS

 

SCREW TERMINAL 

Torque: 0.6 N·m

CHECKING

1) Terminal wiring:  Check that all cables are correctly con-

nected according to the connection diagram.

2) Input:  Check that the input voltage is within 0 – 100% of 

full-scale.

  If the thermocouple or its extension wires are broken, the 

output goes over 100% (below 0% with downscale) due to 

burnout function.  Check leadwires in such a case.

3) Output:  Check that the load is within the permissible 

limit including wiring resistance.  

Load Resistance (Ω) = 

Supply Voltage (V) – 13 (V)

0.02 (A)

(including leadwire resistance)

ADJUSTMENT PROCEDURE

This unit is calibrated at the factory to meet the ordered 
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case 
of regular calibration, adjust the output as explained in the 
following.

 

PREPARATION

1) Loosen Screw (A) on top and open the covering.
2) Remove Screw (B) so that internal adjustment potenti-

ometers are exposed.

A

A

B

C

C

B

3) Remove Screw (C) and take out the circuit boards from 

the housing.

1

2

3

4

5

6

SW3

SW2

VR4

VR5

VR1

VR2

TP1

VR3

SW1

SW1 :  CJC Selector
SW2 :  Zero Bias Selector
SW3 :  Span Gain Selector
VR1  :  0% Adjustment (fine)
VR2  :  100% Adjustment (fine)
VR5  :  0% Adjustment (coarse)
VR4  :  100% Adjustment (coarse)
VR3  :  CJC Sensor Adjustment

6TS1

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201  Fax: +81(6)6659-8510  E-mail: [email protected]

EM-4211  Rev.7  P. 2 / 4