DECIMAL POINT POSITION
Close across the terminals as following.
Use a short-circuit bar or a short wire.
DP1 (×10
1
) LED ON Terminal 7 to 3
DP2 (×10
2
) LED ON Terminal 6 to 3
DP3 (×10
3
) LED ON Terminal 5 to 3
10
3
10
2
10
1
DP3
DP1
10
0
DP2
DISPLAY HOLD COMMAND
Connect the contacts across HOLD to COM.
L
HOLD
H
L
HOLD
H
Twisted pair cable
Twisted pair cable
Relay
Photo-Coupler
COM
HOLD
COM
HOLD
(a) Photo-Coupler
(b) Relay
MAINTENANCE
Regular calibration procedure is explained below:
■
CALIBRATION
1) Turn on the power supply on the unit and keep on sup-
plying the voltage for 10 minutes or more.
2) Remove the front panel.
3) Make a line between terminals 1 to 2 short-circuited and
confirm that the display becomes “000.”
4) Connect a signal generator to the unit, adjust the signal
to the values in the following table and confirm the dis-
plays.
MODEL
SIGNAL VALUE
DISPLAY VALUE
40LV-V1-R
199.0 mV
1990
40LV-V2-R
1.990 V
1990
40LV-V3-R
19.90 V
1990
40LV-V4-R
199.0 V
1990
40LV-A1-R
199.0 µA
1990
40LV-A2-R
1.990 mA
1990
40LV-A3-R
19.90 mA
1990
40LV-A4-R
199.0 mA
1990
40LV-A5-R
1.990 A
1990
5) If the display value is different from the signal value, ad-
just it turning the span adjuster at “1990.”
6) Mount the front panel to the unit again and confirm that
the measurement accuracy is according to the range.