background image

Lyman

®

 Pro-Touch

TM

 1500

Electronic Reloading Scale

110/230 Volt Model Item # 7750718

with Universal Plug Adapters

Introduction

Congratulations on your purchase of the Lyman Pro-Touch 1500 Precision 

Digital scale. This is a precision piece of equipment and should be handled 

with care. Before using the Pro-Touch, please read these instructions fully 

and become familiar with its operation and features. If you do not under-

stand some of the operations or have questions about the use of the scale, 

call us at 1-800-22-LYMAN (Foreign Customers may call 860-632-2020).

Display

The scale is equipped with a backlit, touch screen display. Touch screen 

keys include: ON/OFF which turns the scale on and off, ZERO which tares 

or zeros the scale, MODE which switches the scale from grains to grams, 

and CAL which is used to calibrate the scale.

Set-Up

1.  Select a location for the Pro-Touch that is level and stable. It is recom-

mended that the scale be located away from a loading bench equipped 

with a reloading press. Most benches are not heavy enough to prevent 

vibrations being transferred to the scale. The scale should also be posi-

tioned away from any source of drafts, air currents, or temperature chang-

es. The 1500 is sensitive enough to be affected by these conditions.

2.  Plug the power adapter into a wall outlet and connect it to the scale. The 

scale can also be run on three AAA batteries. To install the batteries, 

remove the cover on the bottom of the scale and install the batteries so 

that the negative ends are against the springs.

3.  The scale is equipped with a clear plastic dust cover. The cover is lift-

ed off when the scale is in use. It is recommended that the cover be 

replaced when the scale is not in use to protect the platform and load 

cell.

4.  For the first use, it is recommended to turn on the scale and allow it to 

run for 24 hours. This will help it to stabilize to its surroundings. 

  Notice: When first turning on the scale it will remember whether 

it was last run in the Grain or Gram mode, and lock in this 

  setting the next time it is turned on. Allow one minute run time 

before changing mode settings.

Calibration

1.  It is recommended that the scale be calibrated before each use, or if the 

scale is moved or jarred in any way.

2.  Press the ON/OFF button to turn on the scale.
3.  Press the ZERO button to reset the scale to zero.
4.  Press the CAL button and the display will read “CAL”, then flash “100.00”. 

Place the supplied 100 gram calibration weight on the platform. The pow-

der pan should not be on the scale.

5.  The scale should read “PASS”, then will show 1543.2 gn on the display. 

Remove the 100-gram weight. The display will now show 0.0 gn.

Operation

1.  Place the empty powder pan on the platform and press the ZERO button 

to reset the scale to zero.

2.  The scale is now ready to weigh powder.
3.  The scale can be changed to display grains or grams as desired. Press 

the MODE button to switch the scale from one to the other. In the grain 

mode the scale will display “grain”, in the gram mode the scale will display 

“gram”.

4.  It is recommended that the ZERO button be pressed on occasion during 

use (approximately every 10 loads measured) to ensure that the scale 

is maintaining its zero or any time the display is off zero when an empty 

powder pan is on the platform.

5.  The scale is equipped with an auto-shutoff feature when batteries are in 

use. The scale will auto-shutoff after approximately 4 minutes of non-use 

to save battery life.

6.  The display will show “LO” to indicate a low battery condition.
7.  To shut-off the scale, press the ON/OFF button.

Cautions

1.  Never leave a load on the scale for a prolonged period of time as this can 

damage the load cell.

2.  Always place the calibration weight or a load gently onto the scale. 

Dropping the weight on may cause permanent damage to the load cell.

3.  Do not attempt to disassemble the scale. There are no user serviceable 

parts inside the scale.

Specifications

Power Supply   . . . . . . . . . . . . . .6 volt DC power adapter or 3 AAA batteries.
Units   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .grains or grams
Capacity/Accuracy . . . . . .1543.2 or - 0.1 or 100.0 grams/+ or - 0.01