(1) If either of the Vertical Shafts detects an abnormal situation the Manual 

Controller will display an error message on the digital display (See Notes   
below  for  error  messages).    Depress  the  “DOWN”  key  for  5  seconds.   
The Vertical Shaft will descend.    Depress the “DOWN” button again to 
complete the reset process. 

 
 
    (2) If during the setup process the digital display shows 560mm but either 
one  or  both  of  the  vertical  shafts  have  not  fully  compressed  depress  the 
“DOWN”  button  for  5  seconds.    The  Vertical  Shaft(s)  will  descend  and  the 
reset operation will be completed.     
 
Notes 
(1) When M1 breaks down, the digital tube displays “ER1”; 
(2) When M2 breaks down, the digital tube displays “ER2”; 
(3)  When  the  temperature  of  the  ELB-A  Controller  exceeds  75

o

C,  the  digital 

display “HOT”.    When power is on, the ELB-A Controller will work normally 
unless the    temperature falls to 45

o

C. 

(4) Working System Protection: The system will operate for 3 to 4 times up and 

down continuously.    After such time the ELB-A Controllerinvokes its inbuilt 
protection circuit and the digital display will show “HOT”.    After 20 minutes, 
the ELB-A Controller will automatically reset and unit will function normally. 

(5) Overcurrent Protection: If the total load of the push rods exceeds 900N, the 

ELB-A Controller will implement the overcurrent protection circuit, and the 
digital diplay will show ER1 or ER2. The load should be reduced and the 
manual reset function used to return the unit to normal functionality. 

 

Picture 2 

 

[Attentions] 

1. This product shall be used in accordance with the stated technology 
parameters; 
2

Keep the unit away from corrosive gas, dusty environments. And use the 

unit in a moisture free environment. 