12-10 Set DO height ( feet ) compensation value
When the lower display show " Highf "
1)This function only for the DO ( Disolved oxygen ) mode of
adjusting the probe's height compensation value in
feet unit. The default value is 0 FEET.
2)Use the " Button " ( 3-5, Fig. 1 ) or " Button "
▲
▼
( 3-6, Fig. 1 ) to select the upper value to the desired
height compensation value ( feet ), then press the
" Enter Button " ( 3-4, Fig. 1 ) will save the setting
value temporally.
12-11 Set CD temperature compensation factor
When the lower display show " PEr C "
1)This function only for the Conductivity ( TDS ) mode of
adjusting the probe's Temp.compensation value in
%/per unit. The default value is 2 %/ per .
℃
℃
2)Use the " Button " ( 3-5, Fig. 1 ) or " Button "
▲
▼
( 3-6, Fig. 1 ) to select the upper value to the desired
Temp. compensation value ( %/per ), then press the
℃
" Enter Button " ( 3-4, Fig. 1 ) will save the setting
value temporally.
12-12 Set CD to TDS or TDS to CD, CD only
When the lower display show " tdS "
1)This function only for the Conductivity ( TDS ) mode to
adjust the function from conductivity ( uS, mS )
to TDS ( PPM ) or from TDS ( PPM ) to Conductivity ( uS,
mS ).