Lutron Electronics MO-2014 Operation Manual Download Page 10

press the " BEEP /       " buttons (3-8 , Fig. 1)  set
warning value , When user to move 1st digit unit  to
next digit unit (10th, 100th. 1000th).

 * Hypothesize display be set at 180.00 .

2) LO warning value setup

Finished (1) procedure then press the " SET / TEST "
button ( 3-4, Fig. 1 ) again, The last digit should flash
at " LO warning value "  on the display . Other
operate way same above explain.

* Hypothesize display be set at 179.00   .

3)Finished (1)(2) procedure then press the " SET /

TEST " button ( 3-4, Fig. 1 ) once . The function
should return to TEST MODE  and indicate measure
value when clipping the resistor. 

4)When measuring period press " COMP /   " ( 3-9 ,

Fig. 1) button  once, The display will show " GOOD "
or " FAIL " symbol to inform be measure results.

For example
a. The value was readout 179.50  , The display

should show " GOOD " symbol and buzzer sounds.(
If buzzer function have been press on status, The
buzzer sounds can to  assist Q.C judgment ) The
resistance value within the accuracy Pass the Q.C.

* Press " BEEP /       " ( 3-8, Fig 1) button once

will start buzzer, press once again to disable 

" BEEP " function.

b. The value was readout 180.10 or 178.60, The

display should show   " FAIL " symbol, The
resistance value out of the accuracy.

 

8

Summary of Contents for MO-2014

Page 1: ...field of p r e c i s i o n measurement A l t h o u g h t h i s Micro Ohm METER is a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques ar...

Page 2: ...ANEL DESCRIPTION 3 4 BASIC 4 WIRES MEASURING PRINCIPLE 4 5 PRECAUTION PREPARATIONS FOR MEASUREMENT 5 6 MEASURING PROCEDURES 5 6 1 Buttons instructions 5 6 2 Symbols units of display 6 6 3 Resistance M...

Page 3: ...C judgment Especial CALCULATE function to measure cable wire length Wide measuring range 5 u 6K 6 ranges Large size and back light LCM display easy read out LSI circuit provides high accuracy reliabi...

Page 4: ...data recorder DL 9602SD 2 2 Electrical Specifications 23 5 Range Resolution Test current Accuracy 60000 u 1 u 10 A 600 m 10 u 1 A 6000 m 100 u 100 mA 0 25 25 d 60000 m 1000 u 10 mA 600 10 m 1 mA 6000...

Page 5: ...3 4 SET TEST key button 3 14 Force Terminal 3 5 HOLD key button 3 15 RS232 socket 3 6 START STOP key button 3 16 Reset button 3 7 L key button 3 17 DC 9V power adapter socket 3 8 Beep key button 3 18...

Page 6: ...WIRES MEASURING PRINCIPAL to maintain the meter in high accuracy Fig 2 Please refer to 2 2 Electrical Specification page 2 each range exists fixed test current from 6000 to 60000 u The fixed current...

Page 7: ...on and Button off ZERO Button When the display indicate few reading before measure SET TEST Press this key to do HI LO limit Button setup Press this key return to TEST status HOLD Button Press this k...

Page 8: ...E mode H I Appears to showing H I limit LO Appears to showing LO limit 6 3 Resistance Measurement 1 Power on the instrument by rotate the Power Range rotary switch 3 10 Fig 1 to exit the OFF position...

Page 9: ...stop measure function and froze the result at LCM 5 When your tested material has inductive you can press L key 3 7 Fig 1 for change to inductor mode at 600 m and 60000 u range Display will showing L...

Page 10: ...tion should return to TEST MODE and indicate measure value when clipping the resistor 4 When measuring period press COMP 3 9 Fig 1 button once The display will show GOOD or FAIL symbol to inform be me...

Page 11: ...nd hold the range 3 Press HOLD button 3 5 Fig 1 to release HOLD status 4 Press SET TEST button 3 4 Fig 1 more than three secoends into sample unit mode 5 Then press L button 3 7 Fig 1 or COMP button 3...

Page 12: ...S232 lead with the following connection will be required to link the instrument with the PC serial port Meter PC 9W D Connector Center Pin Pin 4 3 5 mm jack plug Ground shield Pin 2 2 2 K resistor Pin...

Page 13: ...p bit 1 Stop bit 10 BATTERY REPLACEMENT 1 When the left corner of LCD display show it is necessary to replace the battery However in spec measurement may still be made for several hours after low batt...

Reviews: