An RS232 lead with the following connection will
be required to link the instrument with the PC
serial input.

Scale

PC

(3.5 mm jack plug)

(9W 'D" Connector)

Center Pin........................................ Pin 4

 Ground/shield......................................Pin 2

 2.2 K
 resistor

Pin 5

The 16 digit data stream will be displayed in the following 

format :

D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0

Each digit indicate the following status :

D15

Start Word

D14

4

D13

1

D12,  D11

Annunciator for Display

g = 57

oz = 58

count = 30

D10

Polarity

0 = Positive

1 = Negative

D9

Decimal Point(DP), position from right to the left

0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP

D8 to D1

Display reading, D1 = LSD, D8 = MSD

For example : 

If the display reading is 1234, then D8 to D1 is :

00001234

D0

End Word

RS232 FORMAT : 9600, N, 8, 1

Baud rate

9600

Parity

No parity

Data bit no. 8 Data bits

Stop bit

1 Stop bit

11

1109-GM600P

Summary of Contents for DIGITAL BALANCE GM-600P

Page 1: ...E marks a step forward for you into the field of precision measurement Although this DIGITAL BALANCE is a complex and delicate instrument its durable structure developed Please read the following inst...

Page 2: ...B Button 3 3 5 Display 3 3 6 Power Switch 3 3 7 Rubber Pads 3 3 8 Battery Cover Compartment 3 3 9 DC 9V Receptacle 3 3 10 RS232 Output Socket 3 3 11 Unit Switch 3 4 PLATFORM INSTALLATION 4 5 WEIGHT ME...

Page 3: ...cuit permanently even power off High precision counting scale function RS 232 computer interface Accept battery or AC DC adapter power source two way power supply Built in level bubble the adjustable...

Page 4: ...alibration Use 200 g or 400 g standard weight to execute self calibration automatically Operating 0 to 50 32 to 122 Temperature Operating Humidity Less than 80 RH Size Cabinet 250 x 187 x 70 mm Platfo...

Page 5: ...3 FRONT PANEL DESCRIPTION Side view Bottom view Front view Back view Fig 1 3...

Page 6: ...CAL Button function A Button 3 4 Tare Button function B Button 3 5 Display 3 6 Power Switch 3 7 Rubber Pads 3 8 Battery Cover Compartment 3 9 DC 9V Receptacle 3 10 RS232 Output Socket 3 11 Unit Switc...

Page 7: ...position will set to gram unit the right down corner of LCD will show the g marker Slide the Unit Switch to the left position will set to oz unit 3 Turn the Power Switch 3 6 Fig 1 to the On position 1...

Page 8: ...on again 6 SELF CALIBRATION 1 Please prepare the following optional Standard Weight 200 gram standard weight WT 200 optional or 400 gram standard weight 2 Calibration under the 400 gram standard weig...

Page 9: ...ce again display value of 200 00 g will flash 6 times then the calibration procedures are finished completely balance is ready for the measurement accurately Consideration of Self Calibration The inte...

Page 10: ...e Bi Bi sound release the finger away from function A Button then the bottom left corner of LCD will show the COUNT marker At the same time the LCD will show no of sample units from 10 20 50 100 in se...

Page 11: ...splay will flash twice and freeze the set counting no then take away the sample units Now the balance is ready for counting the sample counting units already change to new value Consideration Under th...

Page 12: ...mpartment replace the battery cover 9 DC 9V AC DC ADAPTER OPERATION Adapter is optional accessory 1 The scale will also be operated by the household ACV power source 110 220 240 ACV with a DC 9V Adapt...

Page 13: ...9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicate the following status D15 Start Word D14 4 D13 1 D12 D11 Annunciator for Display g 57 oz 58 count 30 D10 Polarity 0 Positive 1 Negative D9 Decimal Point...

Reviews: