Optional
* 1.413 mS Conductivity Standard
Accessories
Solution................................
CD-14
SD memory card ( 2 GB )
AC to DC 9V adapter.
USB cable, USB-01.
RS232 cable, UPCB-02.
Data Acquisition software,SW-U801-WIN.
2-2 Electrical Specifications (23±5
℃
℃
℃
℃
)
Conductivity ( uS, mS )
Range
Measurement
Resolution
Accuracy
20 uS
0 to 20.00 uS
0.01 uS
200 uS
0 to 200.0 uS
0.1 uS
2 mS
0.2 to 2.000 mS
0.001 mS
±(2% F.S.+1d)
20 mS
2 to 20.00 mS
0.01 mS
* F.S. -
200 mS
20 to 200.0 mS
0.1 mS
full scale
* Temperature Compensation :
Automatic from 0 to 60
℃
( 32 - 140
℉
), with temperature
compensation factor variable between 0 to 5.0% per C.
* The accuracy is specified under measurement value
≦
100 mS.
* mS - milli Simens
* @ 23±5
℃
TDS ( Total Dissolved Solids )
Range
Measurement
Resolution
Accuracy
20 PPM
0 to 13.20 PPM
0.01 PPM
200 PPM
0 to 132 PPM
0.1 PPM
2,000 PPM
132 to 1,320 PPM
1 PPM
±(2% F.S.+1d)
20,000 PPM
1,320 to 13,200 PPM
10 PPM
* F.S. -
200,000 PPM
13,200 to 132,000 PPM
100 PPM
full scale
* Temperature Compensation :
Automatic from 0 to 60
℃
( 32 - 140
℉
), with temperature
compensation factor variable between 0 to 5.0% per
℃
.
* The accuracy is specified under measurement value
≦
66,000 PPM.
* PPM - parts per million
* @ 23±5
℃
4