02001026
Rev 1 ECO 7062
Operation Manual:
LDD-100/150/250-XX-YY
Page 5 of 13
LDD-100/150/250-XX-YY SPECIFICATIONS
XX = Iout
max
YY = Vout
max
XX * YY cannot exceed Pout
max
Model Pout
max
Iout
max
Input Voltage
Size (L x W x H)
LDD-100-XX-YY
100W
Can be configured
from 10A to 80A
100-240VAC
6.75" x 3.63" x 3.25"
17.1 x 9.2 x 8.26 cm
LDD-150-XX-YY
150W
LDD-250-XX-YY*
250W
Auxiliary Outputs
: +5V @0.25A
+15V @0.25A
-15V @0.25A
Maximum compliance voltage determined by maximum rated power
RS-232 Option available
Other outputs available upon request
Input
Voltage:
100-240VAC,
50/60Hz
Power Factor:
>.98
Interface
Connector:
15 Pin “D” Sub Female
Current Program:
0-10V for 0-Max Current
Current Monitor:
0-10V for 0-Max Current
Voltage Monitor:
0-10V for 0-Max Voltage
Performance
Rise/Fall Time:
~600usec (10% to 90% Full Current)
Current Regulation:
0.5% of Maximum output current
Temperature Drift:
0.5% over temperature range after 30 minute warmup (<0.5% in first 30 minutes)
Current Ripple:
<0.5% of maximum output current
Current Overshoot:
<1% of maximum output current
Power Limit:
Limited to maximum power with power fold-back circuit
Environment
Operating Temp:
0 to 40
o
C
Storage:
-20
to
85
o
C
Humidity:
0 to 90% non-condensing
Cooling:
Forced
air
Regulatory
Leakage Current:
<350uA
Approvals:
Medical Safety:
LDD-100/150-XX-YY: UL60601-1, UL60950-1:2001, IEC 60601-1, EN 60601-
1, CAN/CSA C22.2 No. 601.1-M90
LDD-250-XX-YY:
IEC60950-1:2001
(1
st
Edition), EN 60950-1,2001,
UL60950-1, CSA22.2 No. 60950-03
Emissions/Immunity:
FCC 47 CFR Class A Emissions, EN55011:1998 Group 1 Class A Emissions,
EN61000-3-2 Limits for harmonic current emissions, EN 610000303 Flicker,
EN60601-1-2:2001 Electromagnetic emissions and immunity for medical
equipment
Dimensions:
See Figure 2, LDD-100/150/250 Outline Drawing