Application Note 4 Reliable Operation of LSProbe 1.2 Field Probes
Step 1: Only for E2000 connectors, unclip the con-
nector pair and apply an E2000 locking cap.
Step 2: Expose a fresh area of the lint-free fiber op-
tic cleaning platform.
Step 3: Moisten one corner of the cleaning wipe us-
ing the fiber-optic cleaning pen or IPA pipette.
Step 4: Press the fiber end-face flatly against the
moist area of the wipe and slide it into the dry area
of the wipe.
Step 5: Only for E2000 connectors, remove the
E2000 locking cap, re-clip the connector pair after
cleaning both connectors.
Figure 8: FC/ST/E2000 fiber-optic connector wet-to-dry cleaning method
November 23, 2020