background image

8

Lumex

®

AltaDyne AutoFloat

CPR/Cleaning/Care/Storage

13.0 System Cleaning, Care and Storage

13.1 Cleaning

1.

To clean, use soap, water and a clean cloth to wipe down the Control Unit, power cord, hoses and mattress. Do not use

abrasive cleaners on the mattress. Wipe dry with a clean, dry cloth. Note: Blood and other body fluids must be

thoroughly cleaned from all surfaces before applying disinfectants.

2.

Apply any EPA approved disinfectant to the external surfaces of the Control Unit, hoses and mattress. Allow contact

time according to manufacturers instructions. The solution contact time is what makes disinfection effective.

3.

Wipe down the mattress with a clean, dry cloth to remove any excess disinfectant.

4.

Top sheets of mattress may also be laundered between patient uses or as required to maintain good patient hygiene. 

Fill the washing machine with warm water (70 – 140°F or 21 – 60°C). Add one cup of laundry detergent. Place no more

than four top sheets in a single extra large load capacity washing machine. When wash cycle is complete, remove

promptly from machine and ensure all excess water is drained from load. Place dryer on LOWEST heat setting, 

or AIR FLUFF if available until dry. Verify top sheet is completely dry before placing under patient. Do not dry clean or

iron dry.

5.

If individual air cells of mattress become soiled, clean and disinfect as described above or simply replace air cell with 

a clean replacement. Single air cell replacement can be successfully achieved with patient remaining on the mattress. 

Air cells which do not contain active sensors can also be laundered. A plug is available from Lumex to prevent water

from getting inside these during laundering.

WARNING

Disconnect the AC power cord from the wall outlet before attempting to clean the Control Unit. Do not heat or

steam autoclave any component of the system.

WARNING

Do not launder the Claret colored Active Sensor Technology

air cells of the Active Sensor mattress. Damage may

result. If one or more of the Active Sensor Technology

air cells become damaged, they can only be replaced as a

complete set.

12.0 CPR (Cardio Pulmonary Resuscitation)

To deflate mattress for CPR:

1.

Disconnect the dual quick-connect couplings of the mattress hose from the Control Unit. Mattress will deflate 

in approximately 15-20 seconds.

2.

Proceed with CPR procedures.

13.2 Storage

Control Unit

1.

Check the AC power cord and plug for abrasions or excessive wear.

2.

Coil power cord up and place both cord and Control Unit in plastic bag for storage.

Mattress

1.

Check the air manifold of the mattress for kinks or breaks. Replace if necessary.

2.

To deflate, disconnect hose assembly from Control Unit.

3.

Coil the hose assembly loosely and roll the deflated air mattress starting at the head end. Use the straps for

containment. Place in plastic bag for storage.

Summary of Contents for ALTADYNE 753003

Page 1: ...05 AltaDyne AutoFloat Alternating Pressure Management Plus Model 753003 Model 753004 OPERATORS MANUAL Graham Field Health Products 2935 Northeast Parkway Atlanta Georgia 30360 1 800 347 5678 Fax 1 80...

Page 2: ...ded contact the Graham Field Health Products technical service department Domestic Customers Toll Free 1 800 347 5678 Fax 1 800 726 0601 1 0 Warranty The Control Unit is warranted free of defects in m...

Page 3: ...d frame to prevent mattress from sliding and causing patient injury Lumex mattresses are not intended to be AND DO NOT FUNCTION AS a patient fall safety device SIDE RAILS MUST BE USED WITH THE LUMEX M...

Page 4: ...than that of the patient s skin When the patient moves the Top Sheet tends to move relative to the air cells rather than relative to the patient minimizing shearing effects Additionally the quilting...

Page 5: ...l light while AP mode is active Note AP mode is temporarily suspended in max inflate mode 4 SOFT decreases the inflation level of the bed while Control Unit is in manual mode only Pressing once will d...

Page 6: ...ct the dual quick connect couplings of the mattress hoses to the mating connectors located on the side of the Control Unit 3 Check that the hoses are secure by gently pulling Ensure the hoses are not...

Page 7: ...y before resuming the A P operation 10 2 Use With A Static Mattress 1 Press the Mode Select button until the icons adjacent to the MANUAL and FLOAT icons are both illuminated 2 Adjust the inflation le...

Page 8: ...f mattress become soiled clean and disinfect as described above or simply replace air cell with a clean replacement Single air cell replacement can be successfully achieved with patient remaining on t...

Page 9: ...ttress sensor cord to the jack on the side of the unit Reconnect if necessary Verify hose couplings are fully engaged to controller Verify that all 20 air cells are connected to the air manifold withi...

Reviews: