background image

11

6. Storage

1. Pull the CPR tags and make sure all cells are deflated.

2. Lay the mattress out flat and upside down.

3. Protect the air tube coupler by putting it in the inner side of mattress.

4. Roll the mattress from the head end towards the foot end.

5. Foot-end straps can then be stretched around the rolled mattress to prevent unrolling.

7. Maintenance

7.1 General

1. Check main power cord and plug if there is abrasions or excessive wear.

2. Check mattress' cover for signs of wear or damage. Ensure mattress cover and tubes

are stubbed together correctly.

3. Check the air tube connections for any damage.

4. Report any problems or malfunctions to local distributors.

7.2 Air Filter Replacement

The air filter should be checked and dusted periodically

once per month.
1. Two foam type filters are used. One is located on the

back of pump and the other one is at the bottom of the

Pump.

2. The filter is reusable and can be washed gently with a

mild detergent and water.

3. Suggest to replace air filter regularly if the environment

Is dirty. This will help to improve the airflow. Spare

filters are available from your local distributor.

7.3 Fuse Replacement

1. Disconnect the plug from main power when a fuse is to

be replaced.

2. Remove the cover of the fuse holder by means of a

small screwdriver.

3. Insert a new fuse of the correct rating in, and replace

the cover of the fuse holder back. Check the technical

specification for correct fuse rating.

Summary of Contents for 763400A

Page 1: ...pe er ra at to or rs s M Ma an nu ua al l GF0500067REVA05 LowAirLossTherapy MattressReplacementSystem Model 763400A P Pl le ea as se e r re ea ad d t th he e m ma an nu ua al l b be ef fo or re e u us...

Page 2: ...Contents Important Safeguards 2 Introduction 3 Product Description 4 Installation 5 Operation 7 Cleaning 10 Storage 11 Maintenance 11 Troubleshooting 12 Technical Description 13...

Page 3: ...ice center for examination and repair 5 Keep the cord away from heated surfaces 6 Never block the air openings of this product or place it on a soft surface such as a bed or couch where their openings...

Page 4: ...61000 4 2 IEC 61000 4 3 IEC 61000 4 4 IEC 61000 4 5 IEC 61000 4 6 IEC 61000 4 8 IEC 61000 4 11 EN55011 CLASS B EN 61000 3 2 EN 61000 3 3 FOR US AND CANADA ONLY 1 2 Intended Use The system is a true l...

Page 5: ...It also provides excellent ventilation and keeps patients dry from humidity The 10 inch high mattress which consists of 8 inch high air bladders at the top and a layer of 2 foam at the bottom provide...

Page 6: ...bed frame Please note for the foot end Mattress can be secured firmly by fixing the elastic straps at the bottom bed frame NOTE Please cover the mattress with a cotton sheet to avoid direct skin cont...

Page 7: ...er Do not use it for any other purpose 5 Then turn the main power switch ON The power switch is located on the right side panel of the pump unit NOTE Ensure the CPR connectors are sealed completely In...

Page 8: ...r outage during normal use ALARM MUTE Enable and disable the audible alarm feature Press this button to silence the audible alarm The alarm buzzer will sound again after 3 minutes if the pressure has...

Page 9: ...tomatically STATIC For true low air loss therapy mode similar to a regular air mattress with very low air pressure support and true low air loss therapy PULSATION For pulsating and true low air loss t...

Page 10: ...re Indicator is off it indicates the mattress has reached to its preset softness Patient can lie on the mattress NOTE Check to see if the suitable comfort range or pressure setting is selected by slid...

Page 11: ...button on the control panel When power failure situation occurs alarm will last until backup battery is low or it can be turned off by manual operation Once power failure alarm situation is resolved i...

Page 12: ...oblems or malfunctions to local distributors 7 2 Air Filter Replacement The air filter should be checked and dusted periodically once per month 1 Two foam type filters are used One is located on the b...

Page 13: ...oo long Release some air by disconnecting the air tube connector and change to preferred setting Air Mattress is loose Check if all the snap buttons or straps of mattress are all securely fastened Che...

Page 14: ...orage 10 to 90 non condensing Shipping 10 to 90 non densing Atmosphere 700 hPa to 1060 hPa Classification Class II Type BF IPX0 Applied part Air Mattress Not suitable for use in the presence of a flam...

Page 15: ...U S A Corporate Headquarters Graham Field Health Products 2935 Northeast Parkway Atlanta Georgia 30360 telephone 800 347 5678 770 447 1609 fax 800 726 0601 678 291 3232 http www grahamfield com...

Reviews: