background image

3.  Installation and Operating Instructions

3.1  Contents

The CELESTA light engine ships with the following list of standard components: 

1.

The CELESTA light engine, configured with 7 output channels 
(colors) and an output adapter for connection to an SMA-
terminated optical fiber as documented on the certificate of 
conformance (Figure 1). 

2.

A 24 V/9.2 A DC power supply (Lumencor part number 
27-10019). 

3.

A region-specific AC power cord for the power supply (see 
adjacent table). 

4.

RJ45 ethernet cable 

5.

Control key, external gate jumper, and remote 
interlock jumper (Figure 2) 

6.

Quickstart Guide instruction document 
(85-10049) 

The model name, unique 4- or 5- digit serial number and 
certification markings of the light engine are carried on a 
label affixed to the rear panel.  Performance specifications 
for individual light engines are listed on the certificate of 
conformance included with the shipping documents e-
mailed to the customer (Figure 1). 

3.2  Installation

When setting the CELESTA light engine up for use, place 
the unit on a hard surface and avoid blocking or restricting 
airflow at the air inlets (front panel; Figure 3) or exhaust 
ports (rear panel) on the enclosure.  Restricting the airflow 
will cause the unit to operate at elevated temperatures 
and will result in decreased product life and/or premature 
failure.  The CELESTA light engine must be operated in a 
non-condensing environment (dew point <10ºC with 
controlled ambient temperature <30º C).  Thermal overload 
protection is provided by the on-board computer in 
conjunction with an on-board temperature sensor.  If the 
internal temperature registered by the sensor exceeds 50°C 

OR

 the fan rotor is stopped, all light output channels 

CELESTA Light Engine

®

 Instruction Manual

8

AC Power Cords

Region

Part Number

North America

29-10002

Europe

29-10005

United Kingdom

29-10004

Israel

29-10008

Australia/New Zealand 29-10024

Figure 1. 

 Specimen certificate of conformance (C of 

C) for CELESTA light engine.  The C of C identifies the 
color channels installed in the light engine and the 
bandpass filters associated with each channel.  Full 
(100%) power outputs measured at the terminus of 
the optical fiber are recorded in the third column.

lumencor

DRAFT

14940 NW Greenbrier Parkway  •  Beaverton, OR 97006  •  503-213-4269  •  

www.lumencor.com

Performance Measurements For Lumencor Light Engine S/N

16112

Measurement Method

This product was measured using the custom output adapter installed on the light engine. The

output power was directed at a Coherent thermopile sensor head, model number ______, S/N

__________ attached to a Coherent ______________ Power Meter, S/N ___________. The

measurement system was calibrated on ____________. The meter scale was set to average mode

and the meter zeroed for each measurement. Measurements were taken at room temperature and

at initial turn-on.

Certificate of Conformance

n/a

512/30

555/28

635/22

475/28

747.5/11

440/20

406/20

Bandpass Filter

Color

Channel

Teal

n/a

Green

Red

Cyan

NIR

Blue

Violet

n/a

1000

1002

1019

1024

1013

1001

1005

Power (mW)

Mar 16, 2019

Tammy

Inspected By:

Celesta-NII-7F-A1

Lumencor Build Recipe:

Lumencor Part Number:

Customer Specification:

Document Part Number:  50-10027 rev C

Mar 26, 2018

0405I14R

0615I14R

7064

Total Power:

FieldMaxII-TO

PM30

Customer Part Number:

Fiber 1.5 - A

COC Output Adapter:

Burn-in Pattern

Note:  If this light engine is re-sold, please

ensure that this certifcate is transferred to

the new owner.

n/a

Customer Deviation:

90-10512 Rev. A

SPECIMEN

Summary of Contents for CELESTA Light Engine

Page 1: ...CELESTA Light Engine Instruction Manual Lumencor Inc Document Number 57 10015 www lumencor com ...

Page 2: ...rgy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Safety Certifications TUV SUD America CB Certification IEC 61010 1 2010 TUV SUD America NRTLus Certification UL 61010 1 2012 05 TUV SUD America cNRTL Certification CAN CSA C22 2 No 61010 1 2012 TUV SUD America EN Certification EN 61010 1 2010 Underwriters Laboratories ...

Page 3: ...ecautions and Warnings 3 Installation and Operating Instructions 4 Light Output Characteristics 5 Operational Specifications 6 Routine Maintenance and Troubleshooting 7 Customer Support 8 Warranty CELESTA Light Engine Instruction Manual 3 ...

Page 4: ... The light sources within the CELESTA light engine are controlled by software via one of two serial interfaces USB RS 232 or TCP to an onboard microprocessor operating Lumencor firmware The user can enable or disable each source independently by serial commands as well as change the intensity of each source independently The CELESTA light engine can be controlled by third party microscopy data acq...

Page 5: ...nt des situations ou des pratiques susceptibles d entraîner des blessures corporelles Caution Statements identify conditions or practices that could result in damage to your equipment Attention déclarations qui identifient des situations ou des pratiques susceptibles d endommager le matériel Safety Items Mesures de sécurité Warning ONLY use the power supply provided by Lumencor The Lumencor suppli...

Page 6: ...urité dans un chemin optique fermé NE PAS pointer la sortie de lumière directement sur un matériau susceptible d être inflammable ou susceptible de brûler Cela comprend tous les tissus les plastiques les tissus le papier et les liquides animaux ou végétaux Caution Use of controls or adjustments or performance of procedures other than this specified herein may result in hazardous radiation exposure...

Page 7: ...e parts inside and opening the light engine enclosure will void the manufacturer s warranty Avertissement NE PAS ouvrir l appareil Il ne contient aucune pièce réparable et l ouverture de son boîtier a pour effet d annuler la garantie Caution DO NOT set liquids on the light engine Spilled liquids may damage your light engine Attention NE PAS placer de liquide sur la source lumineuse Les liquides re...

Page 8: ... OR the fan rotor is stopped all light output channels CELESTA Light Engine Instruction Manual 8 AC Power Cords Region Part Number North America 29 10002 Europe 29 10005 United Kingdom 29 10004 Israel 29 10008 Australia New Zealand 29 10024 Figure 1 Specimen certificate of conformance C of C for CELESTA light engine The C of C identifies the color channels installed in the light engine and the ban...

Page 9: ...atus indicator display Figure 3 provides a warning indication that one or more laser sources are active and emitting invisible and or visible radiation The Remote Interlock Connector Figure 2 is provided to allow for connection of a remote interlock When this interlock is open it will shut off laser output After the interlock has been opened the Manual Interlock Reset button will need to be pushed...

Page 10: ...If the light engine is equipped with an output adapter for an SMA terminated optical fiber connect the fiber to the output adapter using the integral threaded sleeve 2 For subsequent start ups use the master power button to start or shut down the light engine Shut down can also be accomplished using the Shut Down button in the Control GUI Figure 4 3 3 3 Ethernet Connection and Control GUI The Web ...

Page 11: ...ference Document Number 57 10018 These commands can be delivered using using Transmission Control Protocol TCP to the LAN input The IP address of the light engine required for LAN communication is shown on the front panel status display Figure 3 or can be obtained using the GET IP command addressed to the RS 232 or USB serial ports The light engine software commands can also be delivered via the U...

Page 12: ...escribed in Section 3 3 3 3 Go to the SETTINGS page in the Web GUI Make sure that the USB port configuration is set to LEGACY mode and USB 5V is set to ENABLED 4 The pod must be set in CELESTA control mode The light engine control mode setting is shown in green letters at the bottom of the pod display screen If the pod is not in CELESTA control mode change the setting by holding down the MODE butt...

Page 13: ... consumption is tracked by the onboard computer If a set limit see Governor Figure 5 is exceeded either by increasing intensity settings for sources that are already on or by turning on additional sources commands will be rejected To clear the error condition reduce intensities of sources that are on or turn off additional sources Light source outputs are refined by bandpass filters The specificat...

Page 14: ... Temperature Operating 32 to 86 F 0 to 30 C Non operating 4 to 158 F 20 to 70 C Humidity Operating and non operating 0 to 80 relative humidity non condensing Altitude Operating 0 to 10 000 feet 3 048 meters Non operating 0 to 45 000 feet 13 176 meters Dimensions Size W x L x H 145 mm x 340 mm x 203 mm 5 7 in x 13 4 in x 8 0 in Weight 8 7 kg 19 1 lbs Lifetime Time for light engine output to decreas...

Page 15: ... all serial ON OFF controls are in the OFF state No light output in response to source ON command serial or TTL There are 4 interlocks that need to be closed check all of these 1 Control key must be inserted in front panel and turned to on position 2 Liquid light guide must be inserted in light output receptacle 3 Remote interlock jumper must be inserted in rear panel 4 External gate jumper must b...

Page 16: ...mance The light engine must also have been used and maintained under operating conditions consistent with the specifications given in Section 5 and observing all the Precautions and Warnings notified in Section 2 This warranty does not extend to light engines that have been subject to misuse accident tampering or improper installation Accessories including but not limited to optical fibers collima...