Input parameters:
- resistance of voltage input [V]
> 1 M
W
- resistance of current input [mA]
12
W
±1%
Sustained overload:
-
TC and RTD
1.1 X
n
-
voltage, current and resistance
1.3 X
n
Short duration overload:
-
voltage input
5 U
n
-
current input
10 I
n
Ensured protection level
acc. to EN 60529:
- housing
IP 40
- from terminal side
IP 20
Dimensions:
22.5
´
100
´
120 mm
Weight:
0.125 kg
Fitting
on a 35 mm DIN rail,
acc. to EN 60715
Electromagnetic compatibility:
- noise immunity
EN 61000-6-2
- noise emission
EN 61000-6-4
Safety requirements acc. to EN 61010-1:
- installation category
III
- pollution degree
2
- phase-to-earth working voltage:
- supply
300 V
2)
- input
50 V
- output
50 V
- altitude above sea level
< 2000 m
1)
A part of sub-ranges for thermocouples and RTD has a specified individual
class – see table 3
2)
Execution for supply voltage 230 V.