Lumax LX-1371 Quick Start Manual Download Page 3

Calibration Procedure: 

The Lumax Electronic Flow Meter permits making quick and precise electronic calibration by changing the calibration factor (k factor). There are 2 dif

ferent ways of calibration:

1.

   On-site calibration, performed by means of a dispensing operation. 

2. 

 Direct calibration, performed by directly changing the k factor.

To enter the calibration phases it is necessary to press and hold down the “Menu” button. 

Why enter the calibration phases?

•   Display the currently used calibration factor 

•   Return to factory k factor after a previous calibration with user k factor

•   Change the calibration factor using one of the two previously indicated procedures.

In calibration mode, the partial and total dispensed quantities indicated on the display screen take on different meanings according to the calibration procedure phase. During the calibration, the Flow 

Meter cannot perform any normal dispensing operations. In calibration mode, the totals are not increased.

Caution:

 The Lumax Electronic Flow Meter features a non-volatile memory. It keeps the calibration and dispensing 

data stored even after replacing new batteries or long periods of inactivity.

3.  Display of Current “K Factor” and Restoring “Factory K Factor”: 

By pressing the MENU key while the appliance is in standby (Press Button for 3 Seconds), 

the display page appears showing the current calibration factor used. If you are using the Flow Meter with 

“factory k factor”, the display page shown in the diagram will be displayed, with the word “fact” 

(Fig. 14)

If one “user k factor” has been set, the calibration factor set by the user (in our example 0.998) will be displayed. 

The word “user” indicates a calibration factor set by the user is being used 

(Fig. 15)

.

The Flow chart alongside shows the switchover logic from one display page to another. In this condition, 

the Reset key permits switching from “User” factor to “Factory” factor. 

To confirm the choice of calibration 

factor, quickly press MENU while “User” or “Fact” are displayed. After the restart cycle, the meter uses the 

calibration factor that has just been confirmed.

Attention:

 When the “Factory Factor” is confirmed, the old “User Factor” is deleted from the memory.

In-field Calibration:

 This procedure calls for the fluid to be dispensed into a graduated sample

container in real operating conditions (Flow Rate, Viscosity, etc.) requiring absolute precision.

WARNING: 

For correct Flow Meter calibration, it is most important to:

•   Completely eliminate air from the system before calibrating

•   Use a precise Sample Container with a capacity of not less than 5 liters, featuring an accurate 

  Graduated indicator.

•   Ensure calibration dispensing is done at a constant flow rate equivalent to that of normal use, 

  until the container is full

•  Not reduce the flow rate to reach the graduated area of the container during the final dispensing 

  stage (the correct method during the final stages of sample container filling consists in making 

  short top-ups at normal operation flow rate)

•   After dispensing, wait a few minutes to make sure any air bubbles are eliminated from the sample 

  container; only read the Real value at the end of this stage, during which the level in the container could drop.

•  If necessary, carefully follow the procedure indicated below.

Fig. 14

1.000

FACT

c

al

Fig. 15

0.998

user

c

al

12.345

Q

TS

23412.3

TOTAL

G

AL

c

al

1.000

FACT

c

al

12.345

USER

c

al

1.000

FACT

12.345

Q

TS

23412.3

TOTAL

G

AL

R
R

C+C

C

STAND BY

STAND BY

C

R+R

Long RESET

R

Short RESET

C+C

Long CAL

C

Short CAL
Time Out

LEGEND:

In-field Calibration Procedure

 Action 

Display

1 NONE

 

Flow Meter in “Stand By” (not in counting mode)

2  LONG MENU KEY KEYING

 

Flow Meter enters calibration mode, shows ”MENU” 

 

and displays the calibration factor in use instead 

 

of total. The words “Fact” and “USER” indicate 

 

which of the two factors is currently in use.

3

 

LONG RESET KEY KEYING

 

Flow Meter shows “MENU” and the partial at zero. 

 

Flow Meter is ready to perform on-site calibration.

4

 

DISPENSING INTO SAMPLE CONTAINER

 

Without pressing any KEY, start dispensing into the sample container.

 

Dispensing can be interrupted and started again at will. Continue dispensing until the  

 

level of the fluid in the sample container has reached the graduated area. There is no  

 

need to reach a preset quantity.

5

 

SHORT RESET KEY KEYING

 

Flow Meter is informed that the calibration dispensing 

 

operation is finished. Make sure dispensing is correctly 

 

finished before performing this operation. To calibrate 

 

the Flow Meter, the value indicated by the partial totalizer

 

(example 9.800) must be forced to the real value marked 

 

on the graduated sample container. In the bottom left part of the display an

 

arrow appears (upwards and downwards), that shows the direction (increase

 

or decrease) of the “USER” K Factor value change when the operations 6 or 7

 

are performed.

 Action 

Display

6  SHORT RESET KEY KEYING

 

Arrow direction changes. The operation can be repeated 

 

if necessary.

7  SHORT/LONG CAL KEY KEYING

 

The indicated value changes in the direction indicated 

 

by the arrow - one unit for every short ”MENU” key 

 

keying - continually if the ”MENU” key is kept pressed 

 

(for the first 5 units slowly and then quickly). If the desired

 

 value is exceeded, repeat the operations from point (6).

8  LONG RESET KEY KEYING

 

Flow Meter is informed that the calibration procedure is 

 

finished. Before doing this, make sure the Displayed 

 

Factor is the Actual Factor.

 

Flow Meter calculates the new “USER” K Factor. This calculation could require a

 

few seconds, depending on the correction to be made. During this operation 

 

the arrow disappears but the ”MENU” indication remains. If this operation is 

 

performed after operation (5), without changing the indicated value, the “USER”

 

K Factor would be the same as the Factory K Factor, thus it is ignored.

9  NO OPERATION

 

At the end of the calculation, the new “USER” K Factor is shown for 

 

a few seconds, after which the restart cycle is repeated to finally 

 

achieve standby condition. 

ATTENTION:

 From now on, the indicated 

 

factor will become the calibration factor used by the meter and 

 

will continue to remain such even after a battery change.

10  NO OPERATION

 

Flow Meter stores the new calibration factor and is ready 

 

for dispensing, applying the newly defined “USER” K Factor

12.345

Q

TS

12.5

TOTAL

G

AL

1.000

Q

TS

G

AL

(user)

c

al

FACT

0.000

Q

TS

c

al

field

9.800

Q

TS

c

al

field

9.800

Q

TS

c

al

field

9.800

Q

TS

c

al

field

9.800

Q

TS

c

al

field

-----

Q

TS

c

al

END

9.860

Q

TS

c

al

FACT

Indicated Value

Real Value

9.86

9.800

Q

TS

c

al

0.0000

G

AL

Indicated Value

Real Value

9.86

1.015

Q

TS

c

al

END

0.000

Q

TS

1234.5

TOTAL

G

AL

c

al

Reviews: