Lumax LX-1371 Quick Start Manual Download Page 2

4.

  Totalizer Register (6 figures from 0.1 to 99999.9), that can indicate two types of Totals:

 

4.1 

 General 

TOTAL that cannot be reset (TOTAL)

 

4.2

   Resettable TOTAL (Reset TOTAL)

  For Totalizer Register - Press “RESET” Button for 3 Seconds

5.

  Indicator of Total Multiplication Factor ( x10 / x100 )

6.

  Display of type of 

TOTAL, (TOTAL / RESET TOTAL)

7.

  Display of Measurement of Totals: L = Liters;  Gal = Gallons; Qts = Quarts;  Pts = Pints

8.

  Display of Flow Rate: 3-30 GPM (10L-120L/min)

9.

  Display of Unit of Measure of the Subset:  L = Liters;  Gal = Gallons; Qts = Quarts;  Pts = Pints

User Buttons: 

Lumax Electronic Flow Meters features Two Buttons (RESET and MENU) which individually perform 

two main functions and together, other secondary functions. The main functions performed are: 

- “

RESET” Key: Resetting the Partial Register and resettable TOTAL

 (

RESET TOTAL) 

- “

MENU” Key: Entering Calibration Mode. 

Used together (Press both Buttons for 5 Seconds), the two keys permit entering Configuration mode, useful for changing the units of measurements and calibration factor. 

Battery Housing:

 Lumax Electronic Flow Meter uses two standard 1.5V power Batteries (size: AAA). Battery Housing is easily accessible - simply loosen four screws on the back of the Device. 

Caution: 

Make sure you remove the 

Batteries very slowly and carefully to ensure that any damage to the Wires, Connections, or the LCD Display Card is avoided.

Installation: 

The inlet and outlet for the Flow Meter is 1“ NPT. It can be easily connected to a Pipe or Nozzle. It can be installed in any desired position and is suitable for stationary or mobile applications. 

Do not use a wrench to install the meter. Hand tighten only. 

Tip:

 To extend the life of the device, it is recommended to install a filter on inlet side of the meter.

1.  Daily Use: 

The only operation that is necessary for daily use are, partial and/or resettable total register resetting. Occasionally the meter may be necessary to be configured or calibrated. For this, 

please follow the instructions in this manual. Below, are the two typical display figures displayed in normal operation.

Figure on the left denotes the Partial and the Resettable Total registers. The figure on the right 

denotes Partial and Non-resettable TOTAL. Resettable total amount adds to non-resettable TOTAL

 

amount automatically. This is preset at the manufacturing plant and cannot be modified.

NOTE:

 6 digits are available for Totals, plus two icons x 10 / x100. The increment sequence as follows: 

0.0 - 99999.9 - 999999 - 100000 x 10 - 999999 x 10 - 100000 x 100 - 999999 x 100

Dispensing in Normal Mode: 

Normal mode is the standard dispensing. While the count is 

made, the partial and resettable total are displayed at the same time: Reset Total 

(Fig. 1)

Should one of the keys be accidentally pressed during dispensing, this will have no effect. 

A few seconds after dispensing has ended, on the lower register, the display switches from 

Resettable TOTAL

 to General 

TOTAL: the word Reset above the word TOTAL

 disappears, and 

the 

RESET TOTAL

 is replaced by the General 

TOTAL

 

(Fig. 2)

. This situation is called standby 

and remains stable until the user operates the Flow Meter again.

Partial RESET:

 The Partial Register can be reset by pressing the RESET

 key when the meter 

is in standby, meaning when the display screen shows the word “TOTAL” 

(Fig. 3)

.

After pressing the RESET

 key, during reset, the display screen first shows all the lit-up digits 

and then all the digits that are not lit up 

(Fig. 4)

.

At the end of the process, a display is first shown with the RESET

 Partial and the 

RESET

 Total 

(Fig. 5)

and, after a few moments, the 

RESET TOTAL is replaced by the Non-Resettable Total 

(Fig. 6)

.

Resetting the RESET TOTAL:

 The 

RESET

 

TOTAL

 resetting operation can only be performed after 

resetting the partial register. The 

RESET TOTAL

 can in fact be reset by pressing the 

RESET

 key at 

length while the display screen shows 

RESET TOTAL

 

(Fig. 7)

. Schematically, the steps to be taken are:

1.

   Wait for the display to show Normal Standby display page - with total only displayed 

(Fig. 8)

.

2.

  Press the 

RESET

 key quickly

3.

  The meter starts to reset the Partial

4.

  While the display page showing the RESET TOTAL

 is displayed 

(Fig. 9)

. Press the 

RESET

 key 

  again for at least 1 second

5.

  The display screen again shows all the segments of the display followed by all the 

  switched-off segments and finally shows the display page where the reset 

 “

RESET TOTAL

” 

(Fig. 10)

 is shown.

Dispensing with Flow Rate Mode Display: 

It is possible to dispense fluids, 

displaying at the same time, as shown 

(Fig. 11)

:

* the dispensed partial

* the Flow Rate in [Partial Unit / Minute]

Procedure for entering this Mode:

- wait for the Remote Display to go to Standby, meaning the display screen shows TOTAL

 only

- quickly press the MENU key.

- Start dispensing

The Flow Rate is updated every 0.7 seconds. Consequently, the display could be relatively unstable at lower Flow Rates. The higher the Flow Rate, the more stable the displayed value. 

IMPORTANT 

-

 

The 

Flow Rate is measured with reference to the unit of measurement of the Partial. For this reason, in case of the unit of measurement of the Partial and TOTAL

 being different, as in the example shown 

below, it should be remembered that the indicated Flow Rate relates to the unit of measurement of the partial. In the example shown, the Flow Rate is expressed in Qts/min. The word “Gal” remaining 

alongside the Flow Rate refers to the register of the Totals (Reset or NON Reset) which are again displayed when exiting from the Flow Rate reading mode 

(Fig. 12)

.

To return to “Normal” mode, press the MENU key again. If one of the two keys RESET or MENU is accidentally pressed during the count, this will have no effect.

Important: 

Even though in this mode they are not displayed, both the 

RESET TOTAL

 and the General Total (

TOTAL

) increase. Their value can be checked after dispensing has terminated, returning to 

“Normal” mode, by quickly pressing MENU key.

Partial RESET:

 To reset the Partial Register, finish dispensing and wait for the Remote Display to show a Flow Rate of 0.0 as indicated in the illustration then 

quickly press RESET 

(Fig. 13)

.

2. Calibration
Definitions: Calibration factor or “k factor”:

 Multiplication factor applied by the system to the electrical pulses received, to transform these into measured fluid units.

FACTORY “K FACTOR”:

 Factory-set default factor. It is equal to 1,000. This calibration factor ensures utmost precision in the following operating conditions:

Fluid: 

Diesel Fuel 

Temperature:

 20°C (68°F) 

Flow Rate:

 10-120 litres/min. (2.65-31.75 Gal.)

Even after any changes have been made by the user, the factory k factor can be restored by means of a simple procedure.

USER K FACTOR:

 Customized calibration factor, meaning modified by calibration.

Why Calibrate?: 

When operating close to extreme conditions, such as for instance with fluids close to acceptable range extremes (like Diesel Fuel at Low Temperatures) or in extreme Flow Rate 

conditions (close to minimum or maximum acceptable values), an on-site calibration may be required to suit the real conditions in which the Flow Meter is required to operate.

Fig. 8

Fig. 9

Fig. 7

Fig. 11

Fig. 12

Fig. 10

4

5 6

9

7

1

2
3

8

88,8,8,8

88888,8

RESET

TOTAL

x100

FLOW RATE

Q

TS

P

TS

G

AL

L

G

AL

L

c

al

12,345

Q

TS

12,3

RESET

TOTAL

G

AL

12,345

Q

TS

12,3

 TOTAL

G

AL

Partial Register

GENERAL

TOTAL

 

Register 

(TOTAL)

RESET TOTAL

(Register Total)

 

RESET TOTAL

Register

Fig. 2

12.345

Q

TS

12.3

TOTAL

G

AL

Fig. 3

12.345

Q

TS

TOTAL

G

AL

23412.3

Fig. 1

12.345

Q

TS

12.3

RESET

TOTAL

G

AL

Fig. 4

88,8,8,8

88888.8

RESET

TOTAL

x100

FLOW RATE

Q

TS

P

TS

G

AL

L

G

AL

L

c

al

Fig. 5

0.000

Q

TS

TOTAL

G

AL

23412.3

Fig. 6

23412.3

0.000

Q

TS

RESET

TOTAL

G

AL

23412.3

0.000

Q

TS

RESET

TOTAL

G

AL

23412.3

12.345

Q

TS

12.3

TOTAL

G

AL

2345.67

0.000

Q

TS

RESET

TOTAL

G

AL

0.0

0.000

Q

TS

RESET

TOTAL

G

AL

12.345

12.3

FLOW RATE

L

Q

TS

G

AL

12.345

12.5

FLOW RATE

Fig. 13

L

L

12.345

0.0

FLOW RATE

Reviews: