LumaSMART
®
User Manual
Modbus Protocol Testing
•
96
134
Configuration Relay 15
0 = Disabled
1 = Normally Deenergized
2 = Normally Energized
Int16
Rel
ay
1
5
135
Alarm Set Point for Relay 15
Int16
136
Restore Set Point for Relay 15
Int16
137
Channels Driving Relay 15
MSB = CH16 / LSB = CH1
Int16
Bit Mapped
138
Control Setting for Relay Number 15
2 = OR Logic
1 = AVG Logic
Int16
139
Configuration Relay 16
0 = Disabled
1 = Normally Deenergized
2 = Normally Energized
Int16
Rel
ay
1
6
140
Alarm Set Point for Relay 16
Int16
141
Restore Set Point for Relay 16
Int16
142
Channels Driving Relay 16
MSB = CH16 / LSB = CH1
Int16
Bit Mapped
143
Control Setting for Relay Number 16
2 = OR Logic
1 = AVG Logic
Int16
144 - 147
Reserved for future expansion
148
Temperature Channel Enable
1 = Enabled
0 = Disabled
MSB = CH16 / LSB = CH1
Int16
Bit Mapped
149 - 240
Reserved for future expansion
6.5.4
Modbus Input Register
Table 9: Modbus Input Register
Address
Parameter
Data Type
0-1
Temperature value for Channel 1
Float
CD AB
Ch
an
ne
l
1
2-3
Probe Signal Strength Channel 1
Float
CD AB
4 - 5
Temperature of Channel 2
Float
CD AB
Ch
an
ne
l
2
6 - 7
Probe Signal Strength Channel 2
Float
CD AB