Chapter 2
____________________________________________________________________
_______________________________________________________________________________________
BI7717-11
1409 Multipoint Sampler
LumaSense Technologies A/S
Service Manual
Page 17 of 28
2.2.6
Reset 1409
The
Reset Command
button will close all open Sample valves, set the 3-way valves in Waste
Air and reboot the firmware.
2.2.7
1409 Multipoint Sampler Status:
The
Read Status
button, will give a read out of the status / error log and 12V power status.
The
Clear Status
button will reset the error log data.
Reset:
indicates if the 1409 has been reset
Power fail:
indicates if the 12 Vdc power is outside the specifications.
Job Errors:
Indicates if an unknown interface command has been send to the 1409.
Error Adress:
Indicates if the 1409 firmware watch dog function has rebooted the
Firmware due to a runtime error.
2.3 Closing the program
Press the Esc button to exit the BZ7006 service software.