BI7717-11
1409 Multipoint Sampler
LumaSense Technologies A/S
Service Manual
Page 10 of 28
1.2.3
Back Panel
Fig.1.3 The back panel of the 1409
Sampler:
From 6 to 24 mounting stubs depended on the 1409 version for
connection of tubing to sampling points. Each stub is numbered,
and has a correspondingly numbered lamp on the Front Plate.
When the lamp is lit, it indicates that the corresponding
sampling valve is open.
AC Mains:
A 3-pin connector accepting Power Cable for connection to a
single phase AC mains supply with protective Earth.
Mains Voltage:
Connect 1409 to mains supply with 100-240 Vac, 50/60 Hz.
Outlet to Analyzer:
Mounting stub for connecting the sampler system of the 1409
to the inlet of an INNOVA Gas Monitor via tubing.Usable models
are 1412i, 1314i, 3434i and 3436i. In the following referred to
as “Gas Monitor”. Please contact LumaSense Technologies A/S
for available models.
Waste Air Outlet:
Mounting stub for connecting the external pump to the 1409’s
sampler system.
Interface:
USB 2.0 Standard
Air Inlet:
The Air Filter is used for an external pump to flush Air when it
does not flush a Sample Channel.