IMPAC pyrometers series IN 5 plus
22
•
When entering a reset time interval other than 0, it must also be determined with “max” or “min” if the
maximum or minimum value storage will be activated.
•
If the pyrometer is equipped with a laser targeting light the laser icon is visible. A click on the laser tar-
geting light icon (
) turns the laser targeting light on or off at this point. After approx. two minutes the la-
ser targeting light is switched off automatically.
•
After switching over to “man”, you can decide if the off-set compensation for the ambient temperature
should be activated (also see 9.2.5 Ambient temperature compensation). The corresponding data
field T(Umg) is then activated so that the ambient temperature value can be entered. In the “auto“ mode
position, the device assumes that the ambient temperature of the measured object is equal to that of the
pyrometer.
The open / save button enable to store and recall own pyrometer configurations.
“1 meas.“ shows the current measuring temperature in the pyrometer parameters
window for approx. 1 second.
A click on the “Test” icon opens a window that
allows the direct communication with the py-
rometer via the interface commands (see sec-
tion 14, Data format UPP
®
).
After entering an interface command (00 is the
adjusted address ex works, “ms“ is the com-
mand “reading temperature value“) and a click
on “Send“ the following window is opened:
This window already shows the answer of the
pyrometer in
1
/
10
°. The actual temperature
reading is 325.7°C.
“Len“ indicates the length of the answered data string, incl. Carriage Return (Chr(13)).
In the lower part of the window the connection with the preset baud rate can be checked. Here the command
was send 500 times with 19200 baud. It has taken 4.56 seconds without transmission errors.
10.8
Measurement color bar
This window displays:
•
current temperature, graphically as color bar
and numerically
•
temperature range or adjusted sub range
•
file size and quantity of the measured values
of the current measurement
•
emissivity
ε
•
the internal temperature of the instrument (T
int
)
•
minimum
(T
min
) and maximum values (T
max
)
•
temperature of the limit contacts
The color bar display shows the span of the temperature
range or the adjusted sub-range. Entering temperature
values in the white fields on the left and right side of the
color bar, limits for the color change of the color bar can be
set. These limits can also be changed by moving the small
bar with the PC mouse. The color bar displays tempera-
tures within the two limits in green color, outside the limits
in red color.
The targeting light (
) can be switched
on or off at this point is the pyrometer is
equipped with it.
In addition, there is an input field
ε
for the emissivity in the window. If the emissivity is changed, the tempera-
ture change connected with this can be read off directly.