INFRATHERM 

pyrometer IS 50-LO/GL 

 

 

15

2 Technical 

data 

 
Temperature range: 

600 ... 1800°C 

Sub-temperature range: 

User-adjustable (minimum span is  51°C). 

Data handling: 

digital 

 

Spectral range: 

0.8 ... 1.1 µm 

Fiber optic cable: 

Multi-fiber light guide (Ø of 0.6 mm) with stainless steel protective coating  

IR detector: 

Silicon (Si)-based photo diode  

 

Power supply: 

24 V DC ± 25% (18 … 30 V DC), ripple must be less than 50 mV 
power consumption max. 0.6 VA 

Analog output: 

4 ... 20 mA direct current, linear temperature resolution (two wire technique) 

Load: 

Max.: 400 

 (18 V) to 700 

 (24 V) to 1000 

 (30 V) 

 

Parameters: 

Adjustable at the pyrometer: 

Emissivity 

ε

, switchable test output 12 mA 

Also adjustable and readable via the PC interface: 

Sub-temperature range (must be within the limits of the full  
temperature range and must have a minimum span of 51°C). 

 

Emissivity: 

5 … 100% 

set via emissivity control inside the pyrometer 

Response time: 

1 s (set ex works) 

 

Accuracy (23°C): 

< 1500°C: within 0.3% of measured value in °C + 1°C 
> 1500°C: within 0.5% of measured value in °C 

Temperature coefficient: 

0.2°C per °C deviation from 23°C  

Reproducibility: 

0.1% of measured value in °C + 1°C 

 

Ambient temperature: 

Converter housing: 

0 … 70°C 

Fibre optics: 

up to 250°C on optical head’s side 

Storage temperature: 

-20 … 70°C 

Enclosure rating: 

IP65 (DIN 40050) 

Enclosure class: 

Ι

 

(VDE 0411) 

 

Weight of signal processor: 

600 g (approx) 

Dimensions of housing: 

125 mm x 80 mm x 57 mm (without plug), die-cast aluminium 

Dimensions of resolving optic: 

112 mm x 22 mm (l x Ø) 

Operation position: 

any 

 

CE-label: 

According to EU directives about electromagnetic immunity 

 
 

2.1 Dimensions 

Signal processor: 
 

  

 

Optical head: 
 

 

 

all dimensions in mm

Summary of Contents for Impac IS 50-LO/GL

Page 1: ...IS 50 LO GL INFRATHERM Pyrometer Betriebsanleitung User Manual IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry www imaxsun com...

Page 2: ...of additional analyzing devices 17 6 Mechanical installation 17 6 1 Overview 17 6 2 Converter 18 6 3 Fibre optical head 18 6 3 1 Ambient temperature 18 6 3 2 Serial number 18 6 4 Minimum bending radiu...

Page 3: ...C Infrared GmbH have a warranty of two years from the invoice date This warranty covers manufacturing defects and faults which arise during operation only if they are the result of defects caused by I...

Page 4: ...e must be within the limits of the full temperature range and must have a minimum span of 51 C Emissivity 5 100 set via emissivity control inside the pyrometer Response time 1 s set ex works Accuracy...

Page 5: ...in a permanent readable condition 4 1 General Each person working with the pyrometer must have read and understood the user manual before operation Also this has to be done if the person already used...

Page 6: ...n side of the power supply switch board the shield must be open to avoid ground loops 5 1 Connector assignment for the converter Connector 3 pole flange plug series 723 Franz Binder GmbH Co Germany Pi...

Page 7: ...ble The fibre optic connectors are equipped with a special sealing ring to protect the connection against penetra tion of liquids Attention The light guide end of the fibre optic cable as well as the...

Page 8: ...on from the measured object which passes through the lens and is then converted to an electrical signal The measured object may be at any distance from the pyrometer However the farther the measured o...

Page 9: ...erature The emissivity setting of the pyrometer has to be adjusted accordingly Typical emissivity values are listed below Measuring object Emissivity at 0 8 1 1 m molten glass 96 Set up examples at th...

Page 10: ...e 4 20 mA current output of the pyrometer This is especially important if the temperature control of your system i e oven or furnace temperature is dependent on the output signal from the pyrometer 8...

Page 11: ...al head 11 2 Accessories 3 820 650 Connection cable length 5 m 3 820 660 Connection cable length 10 m 3 820 670 Connection cable length 15 m 3 820 680 Connection cable length 30 m On request Other len...

Reviews: