luft XC250 Manual Download Page 9

8

Manual XC250

EN

Configuration mode

1. Press the Left/menu key (13) for approx. 2 seconds.

– The device emits a short beep.

– The CFG symbol (14) is displayed in the upper right corner.

2. Use the Right/back or Left/menu key to choose from the

available settings.

– Please note that some of the settings can only be selected

with a certain measuring mode.

Pabs[hPa]

Alarm

Alt.[ft]

CAL12

CFG

20

14

22

23

22

23

Configuration 
mode

Description

Lamp

Setting brightness.
Available are values between 20 and 100 %.
Switch-off after 30 min unless with setting AL.on.

Pabs[hPa]

Only available when the unit g/kg is active.

Specifying the absolute pressure (value range: 600 and 
1200 hPa).
1. Enter the value using the cross control (2).
2. Confirm with the OK button (12).

Alt.[ft.]

Only available when the unit gr/lb is active.

Specifying the local altitude in feet 
(value range: -999 to 9999 ft).
1. Enter the value using the cross control (2).
2. Confirm with the OK button (12).

CAL1 (TH mode)

Specifying the temperature offset.
1. (value range: -10 to +10 °C or -18 to +18 °F)

Enter the value using the cross control (2).

2. Confirm with the OK button (12).

CAL1 (IR mode)

Specifying the temperature offset.
(value range: -10 to +10 °C or -18 to +18 °F)
Set the degree of emission.
(value range: 0.1 to 1.0)
1. Enter the temperature offset value using the cross 

control (2).

2. Press the OK button (12) for 2 s.
3. Enter the emissivity value using the cross control (2).
4. Press the OK button (12) for 2 s.

CAL2

Only available, when % rH is active.

Specifying the offset for relative humidity 
(value range: +/- 30 % RH).
A negative offset allows no measured values of less 
than 0.00.
1. Enter the value using the cross control (2).
2. Confirm with the OK button (12).

ALARM

Only available when IR DP is active.

Stipulating the upper (Hi) and lower (Lo) threshold value 
for the dew point temperature.
value range: up to 9.9 °C or 9.9 °F
1. Enter the value for Hi using the cross control (2).
2. Press the OK button (12) for 2 s.
3. Enter the value for Lo using the cross control (2).
4. Press the OK button (12) for 2 s.
5. Switch the acoustic and visual signal on or off by use of 

the Up (9) and Down (11) keys.

– If the bars (22) in the upper and lower measured

value display are flashing, the acoustic and visual
signal is activated.

– If the bars (23) in the upper and lower measured

value display are flashing, the acoustic and visual
signal is deactivated.

Summary of Contents for XC250

Page 1: ...p p p p p pou ou ou ou ou ou ou u ou ou ou ur r r r r r r r r r la la la la la la la la la la p p p p p p p p p p pr r r r r r r r r r ci ci ci ci ci ci ci ci ci cisi si si si si si si si si ion on on...

Page 2: ...results Lufft Warranty The warranty is for 12 months Damages caused by incorrect use by untrained people or startup by unauthorised people are excluded from the warranty The device complies with the...

Page 3: ...urface temperature of the measuring object and the dew point temperature within the measurement environment A minimum maximum and average function is available for the direct analysis of the measured...

Page 4: ...Technical data Parameter Value Air temperature NTC Measuring range 20 to 50 C Tolerance 0 2 C 0 to 40 C otherwise 0 4 C Resolution 0 1 C Object temperature Pyrosensor Measuring range 70 to 380 C Toler...

Page 5: ...Personnel qualifications People who use this device must have read and understood the operating manual especially the Safety chapter For maintenance or repair work which requires the housing to be ope...

Page 6: ...bag to transport the device safely Storage When the device is not being used observe the following storage conditions dry protected from dust and direct sunlight with a plastic cover to protect it fr...

Page 7: ...device is ready for operation The displayed units are based on the settings of the last utilization 4 Select the desired measuring mode 5 Point the device at the area to be measured The measured value...

Page 8: ...measured value display 18 indicates either the humidity or the dew point To set the unit see Humidity display configuration on page 10 IR DP mode Risk from laser radiation Class 2 laser radiation Las...

Page 9: ...l 2 2 Confirm with the OK button 12 CAL1 IR mode Specifying the temperature offset value range 10 to 10 C or 18 to 18 F Set the degree of emission value range 0 1 to 1 0 1 Enter the temperature offset...

Page 10: ...e more the surface temperature drops the faster increases the repetition frequency of the acoustic and visual signals Maximum repetition frequency is triggered when reaching the lower alarm limit TdP...

Page 11: ...values cannot be saved on the device itself In order to save measured values the device has to be connected to a PC via a USB cable using the SmartGraph software 1 Briefly press the OK button 12 The...

Page 12: ...adiate Many organic materials and surfaces have a degree of emission of approx 0 95 Metal surfaces or shiny materials have a lower degree of emission and therefore only yield inaccurate measured value...

Page 13: ...isit the website www smartgraph3 de download 2 Double click the downloaded file 3 Follow the instructions of the installation wizard Starting the PC software 1 Start the SmartGraph3 software 2 Switch...

Page 14: ...il of 27th January 2003 concerning old electrical and electronic equipment After the end of its use please dispose of this device in a manner appropriate to the relevant legal requirements In the Euro...

Page 15: ...820 E Mason St A Santa Barbara CA 93103 Tel 01 919 556 0818 Fax 01 805 845 4275 E Mail sales lufftusainc com www lufft com Lufft China Shanghai Of ce Lufft Shanghai Measurement Control Technology Co...

Reviews: