Operating Manual CHM 8k
R1.7 / 01-2020
Installation
19
Figure 10 Schematic sketch of the electrical installation.
1 Power supply (3-pole)
2 Mains switch and electric junction box
3 Data transfer cables: RS485 or LAN,
4 Data cable connection: RS485 or LAN
5 PC for remote access (via LAN/DSL; PC must not be local)
6 Ground connection ceilometer
7 Lightning conductor
WARNING
Touching live parts may result in electric shock, which will cause serious or fatal
injury.
Switch off the external circuit breaker before starting the installation and secure
against reconnection.
The electrical connections to the CHM 8k are specified in more detail in Figure 11. Connect the mains
and data cables as shown in this figure. It is recommended to include external overvoltage protection in
all connections to the device to prevent damage to the connection box. Internal lightning protection is
provided in the device.
The
CHM 8k is connected via the following supplied cables: