Operating Manual
ARS31(Pro)-UMB
G. Lufft Mess- und Regeltechnik GmbH, Fellbach, Germany
32
10 Troubleshooting
Description
Cause / remedy
The sensor will not allow
polling requests.
-
Sensor (no dummy cover) installed
-
Check power supply voltage in the connection
(terminal) box and on the sensor
-
Check interface connection
-
False device ID
check ID
-
RS485 cable A / B reversed
-
False baud rate set (sensor, ISOCON-UMB)
-
False protocol set or used (binary, ASCII)
Freezing temperature is
not calculated.
-
Start conditions not fulfilled (road temperature above
set temperature)
-
Sensor surface dry
-
Measurement cycle not yet completed (lasts up to 20
minutes)
-
Interference during calculation of freezing temperature
due to external influences in critical phases of the
measurement cycle
-
Sensor surface heavily soiled -> clean sensor surface
with water
-
Green protective foil not removed ->
remove green
protective foil
Determined freezing
temperature is reset after
a few minutes
-
Sensor is resetting
-
Power supply overloaded
-
Unstable power supply
-
Related to an LCOM-UMB: Misconfiguration of LCOM,
thereby resetting the UMB-Modules every x minutes
(default is 15 Minutes)
Salt concentration channel
delivers error message
-
Start conditions not fulfilled (road temperature above
set temperature)
-
Sensor surface dry
-
Measurement cycle not yet completed (lasts up to 20
minutes)
-
Interference during calculation of freezing temperature
due to external influences in critical phases of the
measurement cycle
-
Sensor surface heavily soiled ->
clean sensor surface
with water
-
Green protective foil not removed->
remove green
protective foil
The description of the UMB-
errorcodes you can find in the document “UMB-Protocol” under
the sector Support
– Downloads – Manuals – UMB Technology.