Chapter 6 Equipment Description
9
6.4.1
Heating mode
The heating of the device can be operated in 4 different modes:
off:
Heating always off
auto:
Automatic heating control
1
The heating switches on when the housing temperature falls below +15°C
(adjustable between 2°C and 20°C) and switches off at a housing temperature of
>+20°C (set tempe5°C)
(metal)
The heating switches on when the housing temperature falls below +50°C
(adjustable between 2°C and 70°C) and switches off at a housing temperature of
>+55°C (set tempe5°C)
(plastics)
testmode:
The switch-on temperature is adjusted to +40°C; in this condition the heating
switches on at room temperature (for test purposes only)
ctrl high:
Heating control is disabled when the control line is at the “high” level, else
automatic
ctrl low:
Heating control is disabled when the control line is at the “low” level, else automatic
For configuration of the heating mode see chapter 12.
6.4.2
Heating capacity
The heating capacity can be set in accordance with the following modes:
full:
Full heating capacity (ca. 240W)
1
alternating:
Alternating heating:
Cover plate alternating to base plate (ca. 100W or 150W alternating)
The next level of heating switches in if the pre-set temperature for the
respective level is not reached within 4 minutes.
In this mode you only need a power supply with 140W.
For configuration of the heating capacity see chapter 12.
6.4.3
Heater Turn Off on Over / Under Voltage
If the heater turn off on over / under voltage is activated, the heating will be automatically
switched off, if the supply voltage is outside of the surveillance window.
The lower threshold is ~20V, the upper threshold is ~28V.
For configuration of the turn-off-function see chapter 12.
1
Factory setting