background image

Operating Manual

 

UMB Analogue/Digital Converter ANACON 

 

G. Lufft Mess- und Regeltechnik GmbH, Fellbach, Germany

 

 

1

 

 

 

Operating Manual 

UMB Analogue/Digital Converter ANACON

 

Order Number: 8160.UANA

 

Status: V6; 1425.03.2010 

                      

 

 

                                              

 

 

Summary of Contents for 8160.UANA

Page 1: ...ting Manual UMB Analogue Digital Converter ANACON G Lufft Mess und Regeltechnik GmbH Fellbach Germany 1 Operating Manual UMB Analogue Digital Converter ANACON Order Number 8160 UANA Status V6 1425 03...

Page 2: ...UFACTURER 20 EC CERTIFICATE OF CONFORMITY 21 Version history Version Date Compiled by Comments V1 15 09 2006 EES SH First edition V2 11 01 2007 EES Sensor connection scheme for wind sensor changed add...

Page 3: ...witched off Protect the device from direct sunlight The equipment requires an adjustment period of several minutes on being moved to a location with different climate conditions Pay attention to the t...

Page 4: ...nverter ANACON Operating Manual 4 G Lufft Mess und Regeltechnik GmbH Fellbach Germany Symbols used Important indication concerning possible hazards to the user Important indication for the correct fun...

Page 5: ...tion of the ANACON via RS232 port and an additional ISOCON 8160 UISO Features 2 analogue inputs with 24 bit resolution and signal conditioning for resistance PT100 PT1000 voltage current frequency and...

Page 6: ...ital Converter ANACON Operating Manual 6 G Lufft Mess und Regeltechnik GmbH Fellbach Germany ANACON UMB Analogue Digital Converter Summary Status LED s Input Channel B Input Channel A Power supply RS4...

Page 7: ...ACON G Lufft Mess und Regeltechnik GmbH Fellbach Germany 7 Analogue Inputs Channel A and Channel B ANACON UMB Connection Channel B Channel A PIN Name 1 Power Sensor Power Supply 2 ANA_ Sense High Inpu...

Page 8: ...al 8 G Lufft Mess und Regeltechnik GmbH Fellbach Germany Mounting Rail Bus Connector NOTE The assembly must take place on a grounded DIN rail Earth 24V Power Supply Feed Earth 24V Power Supply Feed 24...

Page 9: ...ration and functioning of the ANACON Their meanings are as follows Red LED If this LED lights up permanently after switch on there is a configuration fault If this LED lights up during operation an un...

Page 10: ...se ensure that the polarity and pin assignment are correct on connection Incorrect connection can cause the destruction of the ANACON module 2 Next mount the ANACON module which you wish to configure...

Page 11: ...B of the ANACON module Please note that a fixed assignment is specified for channels A and B for the combined sensors for temperature and relative humidity as well as the wind speed and wind direction...

Page 12: ...4 Resistance 3 wire PT100 PT1000 Voltage V V V V Current self powered I I I I Current ANACON powered I I I I Frequency F F F F Impulse Digital Input D D D D Lufft Sensors 8160 TF T red red white red r...

Page 13: ...50 0 C 105 125 145 165 A temperature 328 0 842 0 F 101 121 141 161 B temperature 200 0 450 0 C 106 126 146 166 B temperature 328 0 842 0 F 110 130 150 170 dewpoint 200 0 450 0 C 111 131 151 171 dewpoi...

Page 14: ...kts 501 521 541 561 581 B Wind Direction 0 359 9 Precipitation 600 A precipitation absol mm 601 B precipitation absol mm 620 A precipitation diff mm 621 B precipitation diff mm 640 A precip intens mm...

Page 15: ...40 10260 A Resistance 0 2000 Ohm 10201 10221 10241 10261 B Resistance 0 2000 Ohm Frequency 10300 10320 10340 10360 A Frequency 10 10000 Hz 10301 10321 10341 10361 B Frequency 10 10000 Hz Pulses 10500...

Page 16: ...ction Phoenix bus connector gold 5 pole Function display Green LED flashes every 10 seconds RS485 interface internal bus Standard RS485 2 wire half duplex Transmission type Protocol transparent 8 data...

Page 17: ...mperature with PT100 200 500 C 0 02 C for 50 C t 150 C otherwise 0 1 C 0 1 C for 50 C t 150 C otherwise 0 5 C Temperature with PT1000 200 500 C 0 02 C for 50 C t 150 C otherwise 0 1 C 0 1 C for 50 C t...

Page 18: ...ure Humidity Sensor 8160 TFF Temperature PT1000 30 70 C Resistance 882 2 1270 8 Ohms Humidity 0 100 RH Frequency Barometric Air Pressure Sensor 8355 03 Air Pressure 0 1200 hPa Current 4 20 mA Wind Spe...

Page 19: ...ent temp 40 70 C Permissible rel humidity 0 95 RH non condensing Operating conditions Permissible operating temp 30 60 C Permissible rel humidity 0 95 RH non condensing Housing Material PA V0 green Wi...

Page 20: ...with European Directives 2002 96 EC and 2003 108 EC waste electrical and electronic equipment Waste equipment must not be disposed of as household waste For environmentally sound recycling and the di...

Page 21: ...d the Low Voltage Directive in accordance with 73 23 EC The above mentioned equipment specifically conforms to the following EMC Standards EN 61000 6 2 2005 Part 6 2 Generic Standards Immunity for ind...

Reviews: