Model 3
Technical Manual
Section 7
Ludlum Measurements, Inc.
Page 7-3
April 2016
SYMPTOM
meter goes full scale
or “pegs out”
(continued)
no response to
radiation
no audio
POSSIBLE SOLUTION
3.
Remove the can and check for loose
or broken wires.
4.
Ensure that the instrument’s “can”
is properly attached. When attached
properly, the speaker will be located
on the left side of the instrument. If
the can is on backwards, interfer-
ence between the speaker and the
input preamplifier may cause noise.
1. Substitute a “known good” detector
and/or cable.
2. Has the correct operating voltage
been set? Refer to the calibration
certificate or detector instruction
manual for correct operating
voltage. If the instrument uses
multiple detectors, confirm that the
high voltage is matched to the
current detector being used.
1. Ensure that the
AUD ON
-
OFF
switch
is in the
ON
position.
2. Remove the instrument housing and
check the connection between the
circuit board and the speaker. Plug
in the 2-pin connector if necessary.
Troubleshooting GM Detectors
1. If the tube has a thin mica window, check for window breakage.
If damage is evident, the tube must be replaced.
2. Check the
HV
. For most GM tubes, the voltage is normally 900
Vdc, or 460-550 Vdc for “peanut” tubes (Ludlum Model 133
series).
Summary of Contents for MODEL 3
Page 1: ...LUDLUM MODEL 3 SURVEY METER April 2016 Serial Number 294605 and Succeeding Serial Numbers ...
Page 2: ...LUDLUM MODEL 3 SURVEY METER April 2016 Serial Number 294605 and Succeeding Serial Numbers ...
Page 4: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......