CONNECTOR: 

series "C" 

(others available)  

CONSTRUCTION: a

nodized aluminum housing   

TEMPERATURE RANGE: 

-15 to 50 °C (5 to 122 °F); may be 

certified to operate from -40 to 65 °C (-40 to 150 °F) 

SIZE: 

4.6 x 14.7 cm (1.8 x 5.8 in.) (Dia x L)                        

WEIGHT: 

0.5 kg (1 lb)   

 

 

PRELIMINARY INSTRUCTIONS 

 

UNPACKING AND REPACKING   

Remove calibration certificate or detector functional check certificate 

and place it in a secure location. Remove the detector and accessories 

(cable, etc.) and ensure that all of the items listed on the packing list are 

in the carton. If more than one detector is in the carton, refer to the 

calibration certificate(s) for serial number  (S/N) match. The Model 

44-7 S/N is located on the side of the detector near the connector. To 

return the instrument for repair or calibration,  provide sufficient 

packing material to prevent damage during shipment and appropriate 

warning labels to ensure careful handling. 

 

The following items and information should also be included to 

ensure a quick turn-around time on your repair calibration:

   

 

instrument(s) and related cable(s)  

 

brief information as to the reason for return   

 

description of service requested   

 

return shipping address   

 

customer name and telephone number   

 

NOTE: When shipping a Model 44-7 by air, it is necessary 

to ship the tube in a  sealed container to avoid sudden 

atmospheric changes, which could rupture the tube.   

Summary of Contents for 44-7

Page 1: ...LUDLUM MODEL 44 7 Alpha Beta Gamma Detector July 2012 Serial No PR090405 and succeeding Serial Numbers ...

Page 2: ......

Page 3: ...LUDLUM MODEL 44 7 Alpha Beta Gamma Detector July 2012 Serial No PR090405 and succeeding Serial Numbers ...

Page 4: ...anty of merchantability or fitness which extend beyond the description of the face there of If the product does not perform as warranted herein purchaser s sole remedy shall be repair or replacement at the option of Ludlum Measurements In no event will Ludlum Measurements be liable for damages lost revenue lost wages or any other incidental or consequential damages arising from the purchase use or...

Page 5: ...Table of Contents General 1 Specifications 1 Preliminary Instructions 2 Operating Procedures 3 Parts List 4 Assembly Drawing 5 Response Curve 6 ...

Page 6: ...ely 900 V The recommended instrument input sensitivity is approximately 30 mV or higher to prevent the detector from double pulsing The GM tube face can rupture above 2438 m 8000 ft altitude pressure Consequently detectors carried in unpressurized aircraft above this altitude would be subject to failure SPECIFICATIONS DETECTOR end window halogen quenched GM WINDOW 1 7 0 3mg cm2 WINDOW AREA active ...

Page 7: ... the carton refer to the calibration certificate s for serial number S N match The Model 44 7 S N is located on the side of the detector near the connector To return the instrument for repair or calibration provide sufficient packing material to prevent damage during shipment and appropriate warning labels to ensure careful handling The following items and information should also be included to en...

Page 8: ... detector 2 Connect the detector to the instrument and check for a proper background reading typically 25 to 50 cpm at a background level of 8 to15 µR hr 3 Expose the detector to a check source and verify that the instrument indicates within 20 of the check source reading obtained during the last calibration Alternatively expose the detector to a check source of a known value and verify that the d...

Page 9: ... 1536 Model 44 7 alpha beta gamma detector 1 ea Detector Body 7002 065 01 1 ea End Cap 7002 065 02 1 ea GM Tube 01 5006 LND 723 TGM N210 2 ea Sponge 7002 065 04 1ea Socket 01 5025 1 ea Connector UG 706 U 13 7751 1 ea O Ring 2 128 16 8284 3 ea Screws 4 40 x 3 16 FH 17 8811 1 ea Red Protective Cap 03 5474 1 ea Cap Model 44 7 Grid with O Ring 4002 216 ...

Page 10: ...Model 44 7 Technical Manual Page 5 July 2012 ...

Page 11: ...Model 44 7 Technical Manual Page 6 July 2012 ...

Reviews: