Model 42-49 Neutron Detector      

Technical Manual 

 

 

Section 5

 

Ludlum Measurements, Inc.

 

Page 5-1 

December 2017 

 

 

Calibration  

Calibration  

To calibrate the detector, the following items are needed: 

 

 

Counting instrument: Ludlum Model 2200 Scaler/Ratemeter 

 

 

241

AmBe neutron source or equivalent 

 

 

137

Cs gamma source for gamma rejection check 

 

Operating Voltage Plateau  

  Connect the Model 42-49 to a Model 2200.  

 

  Set Model 2200 input sensitivity to -2 mV. 

 

  Expose the detector to a 20 mrem/hr 

241

AmBe neutron source.   

  

  “Run  a  plateau”  as  follows:  Taking  one-minute  counts,  record  the 

count  rate  for  25-volt  incremental  steps  from  1000  volts  through  a 
region  in  which  the  count  rate  steadily  increases,  becomes  relatively 

constant  (“flattens  out”),  and  then  increases  again.    (A  typical  range 
will  be  1000-1250  volts.)  The  flattest  portion  of  this  region  is  the 

desired plateau. The value in the middle of this region is the assumed 
operating  voltage  of  the  detector,  subject  to  the  sensitivity  and  the 

gamma rejection checks. 

 

  Calculate  the  sensitivity  (cpm/mrem/hr)  for  the  assumed  operating 

voltage as follows: 

Rate

Equivalent

Dose

Rate

Count

y

Sensitivit

 

Section 

Summary of Contents for 42-49

Page 1: ...LUDLUM MODEL 42 49 NEUTRON DETECTOR December 2017 ...

Page 2: ...LUDLUM MODEL 42 49 NEUTRON DETECTOR December 2017 ...

Page 3: ...Repacking 2 Connecting to an Instrument 2 1 Specifications 3 Safety Considerations 4 Environmental Conditions for Normal Use 4 1 Cleaning Instructions and Precautions 4 1 Calibration 5 Calibration 5 1 Operating Voltage Plateau 5 1 Gamma Rejection Check 5 2 Conversion Chart 5 3 Parts List 6 ...

Page 4: ...ese gamma rays in the detection process A common reaction for the conversion of slow neutrons into directly detectable particles is n 3 He 3 H 1 H 0 764 MeV The Ludlum Model 42 49 utilizes this reaction in the form of helium 3 3 He which fills the gas proportional tube of the detector The Model 42 49 is designed to be used with portable counting instruments and has a top bracket that allows for co...

Page 5: ...lManual Section1 Ludlum Measurements Inc Page 1 2 December 2017 Note The detector does not contain any consumable materials Note If the detector is used in a manner not intended by the manufacturer the detector may not function properly ...

Page 6: ...calibration provide sufficient packing material to prevent damage during shipment Every returned instrument must be accompanied by an Instrument Return Form which can be downloaded from the Ludlum website at www ludlums com Find the form by clicking on the Support tab and selecting Repair and Calibration from the drop down menu Then choose the appropriate Repair and Calibration division where you ...

Page 7: ... mV input sensitivity Sensitivity 100 cpm mrem hr 241 AmBe fast neutrons Gamma Rejection 10 cpm or less through 10 R hr 100 mSv hr 137 Cs Detection Range thermal to approximately 12 MeV Energy Response thermal to 12 MeV follows the radiation protection guide curve for neutron dose Input Sensitivity 2 mV Operating Voltage approximately 1200 V Connector series C others available Temperature Range 20...

Page 8: ...less than 95 non condensing Pollution Degree 2 as defined by IEC 664 Normally only nonconductive pollution occurs Temporary conductivity caused by condensation is to be expected Cleaning Instructions and Precautions The detector may be cleaned externally with a damp cloth using only water as the wetting agent Do not immerse the instrument in any liquid Observe the following precautions when cleani...

Page 9: ...em hr 241 AmBe neutron source Run a plateau as follows Taking one minute counts record the count rate for 25 volt incremental steps from 1000 volts through a region in which the count rate steadily increases becomes relatively constant flattens out and then increases again A typical range will be 1000 1250 volts The flattest portion of this region is the desired plateau The value in the middle of ...

Page 10: ...ly 120 cpm mrem hr Gamma Rejection Check With the Model 42 49 connected to the Model 2200 adjust the Model 2200 HV to the assumed operating voltage determined above Remove the Model 42 49 detector from the moderator and place in a 10 R hr 137 Cs gamma radiation field Take a one minute count If more than 10 counts are observed for the one minute period decrease the operating voltage until the count...

Page 11: ...0 mrem hr Take a one minute count and record the value including the range scale setting of counting instrument Repeat for the dose equivalent rates shown in Table 1 The values in Table 1 and their corresponding measured values represent a conversion chart for use in relating other measured values to actual dose equivalent rates Table 1 Ref Point mrem hr Reading cpm Range Scale 400 200 80 20 8 2 ...

Page 12: ...05 060 7 7 inch SPHERE 7014 858 2 EA INSERT E Z LOK 6 32 20 9066 8 EA 8 INCH HDPE ROD 29 9985 MODEL 42 49 DET MODERATOR 47 4059 1 EA O RING 2 116 16 8293 1 EA O RING 2 148 16 8340 2 EA SCREW 8 32 X 5 8 SOC SET 17 8765 MODEL 42 49 PROBE ASSY 4014 587 3 He TUBE 25185 2 ATM 01 5793 CONNECTOR UG706 U 4478 011 TUBE BODY ASSY 2014 595 MODEL 42 30 42 31 DETECTOR SPACER 7005 056 CONNECTOR END 7014 598 CBL...

Reviews: