Ludlum Measurements 44-110 Manual Download Page 7

 

Model 44-110 & 44-110-1 Large Area Tritium Detectors 

 

 

 

Page 2 

approximately  15  seconds  with  a  gas 
flow  rate  of  7  liters  per  minute.  After 
obtaining the background measurement, 
the  gas  valve  can  be  turned  to 

OFF

 

(Model  44-110)  or 

STANDBY

  (Model 

44-110-1) until the start of the survey. 

 
2. 

For surveying, place the detector on the 
surface  to  be  assessed  and  turn  the  gas 
valve  to 

ON

  or 

NORMAL

.  Observe  the 

count-rate  meter  until  the  maximum 
background count   is   reached. Slowly 
slide  the  detector  along  the  surface, 
pausing for a moment for each detector 
width.  If  any  increase  in  count  rate 
above background is observed, keep the 
detector  over  the  suspected  area  until 
maximum reading is obtained. 

 
3. 

Survey Considerations 

 

a.

 

Using  the  Model  44-110,  very  smooth 

surfaces  such  as  tile  floors  and  lab 
bench  tops  can  be  surveyed  very 
effectively  and  rapidly  because  the 
detector can be purged once and then 
slid over the survey surface area.  

 
 

 

b.

 

The Model 44-110-1 is more suited for 

surveying less regular surfaces and for 
personnel frisking. 
 

c.

 

When  surveying  less  regular  surfaces 

with  the  Model  44-110,  the  probe 
cannot  be  slid  across  the  surface  and 
re-purged for each measurement. 

 

d.

 

Irregular,  curved,  and  very  small 

surfaces,  such  as  drain  pipes,  table 
legs,  etc.,  cannot  be  effectively 
assessed  using  either  of  these 
detectors. 

 

NOTE

 

 

In order to prevent residue forming on the 
chamber  and  anode  wires,  the  detector 
should not be used on very dirty or dusty 
surfaces.  Extreme  humidity  in  the  air  (in 
excess of 85%) and moisture on surfaces 
can  cause  the  detector  to  behave 
erratically. 

Therefore, 

it 

is 

not 

recommended  to  use  the  detector  in high 
humidity or in moist conditions.  
 
Static electricity on a surface is sufficient 
to  prevent  some  tritium  beta  particles 
from  escaping  from  the  surface,  thus 
reducing calculated surface emission.

5. 

MAINTENANCE

 

 

Dust  on  the  anode  wires  may  impair  the 
efficiency.  This  is  manifested  by  changes  in 
the  background  count  rate.  A  background 
count rate change of ±150 cpm may be due to 
dust  accumulation,  rather  than  an  actual 
change 

in 

background 

or 

low-level 

contamination.  A  careful  assessment  is 
required  to  determine  the  exact  cause  of 
variation in count.   
 
Surveying  in  a  dusty  environment  may 
require cleaning the anode wires and chamber 
several times a day. 

 

 
Cleaning anode wires:

  

Use a dust remover spray such as those used 
by  photographers,  followed  by  a  spray  of 
electrical  contact  cleaner.  The  cleaner  or 
solvent must be 100% residue free.   
 

CAUTION 

 

When  cleaning  the  counting  chamber,  be 
extremely careful not to damage or break the 
delicate anode wires. 

Summary of Contents for 44-110

Page 1: ...LUDLUM MODEL 44 110 44 110 1 LARGE AREA TRITIUM DETECTORS March 2013 ...

Page 2: ...LUDLUM MODEL 44 110 44 110 1 LARGE AREA TRITIUM DETECTORS March 2013 ...

Page 3: ...anty of merchantability or fitness which extend beyond the description of the face there of If the product does not perform as warranted herein purchaser s sole remedy shall be repair or replacement at the option of Ludlum Measurements In no event will Ludlum Measurements be liable for damages lost revenue lost wages or any other incidental or consequential damages arising from the purchase use or...

Page 4: ......

Page 5: ...Model 44 110 44 110 1 Large Area Tritium Detectors TABLE OF CONTENTS 1 GENERAL 1 2 SENSITIVITY 1 3 COUNTING GAS 1 4 OPERATION 1 5 MAINTENANCE 2 6 DRAWINGS DIAGRAMS 3 ...

Page 6: ...iciency in a 400 cpm background and a count rate instrument with an approximate 22 second response time is approximately 503 dpm 100 cm2 3 COUNTING GAS Recommend 300 ft3 P 10 cylinder mounted on a cylinder truck for mobile gas supply Estimated time of operation for one 300 ft3 cylinder is approximately 40 hours 4 OPERATION 1 Connect the P 10 counting gas and HV signal cable to the detector Place t...

Page 7: ...ved and very small surfaces such as drain pipes table legs etc cannot be effectively assessed using either of these detectors NOTE In order to prevent residue forming on the chamber and anode wires the detector should not be used on very dirty or dusty surfaces Extreme humidity in the air in excess of 85 and moisture on surfaces can cause the detector to behave erratically Therefore it is not reco...

Page 8: ...NGS AND DIAGRAMS M 44 110 TRITIUM FRISKER Drawing 342 170A M 44 110 ASSY Drawing 342 170B 170D M 44 110 1 VALVE ASSEMBLY Drawing 342 171 M 44 110 1 PLUMBING ASSEMBLY Drawing 342 171A M 44 110 1 VALVE CALIBRATION Drawing 342 171B FACE PLATE M 44 110 1 Drawing 342 172 ...

Page 9: ......

Page 10: ......

Page 11: ......

Page 12: ......

Page 13: ......

Page 14: ......

Page 15: ......

Page 16: ......

Reviews: