Model 44-110 & 44-110-1 Large Area Tritium Detectors 

 

 

 

Page 1 

 

1. 

GENERAL

 

 
 

The  Ludlum  Model  44-110  and  44-110-1 
Tritium  Detectors  are  large  area  windowless 
gas  flow  proportional  detectors  used  for 
detecting fixed tritium contamination. Due to 
the energy of tritium, certain restraints of the 
detector  make  it  marginally  more  difficult  to 
use than many of the other types of detectors.  
 
A  count  rate  instrument  that  is  capable  of 
providing an operating voltage of 1750 V 

±

50 

V and a threshold of 4 mV ±1 mV is required. 
 

 
The  Model  44-110-1  is  different  from  the 
Model  44-110  in  that  it  is  equipped  with  a 
multi-position  gas  flow  selector  valve.  The 
selections  available  by rotary  switch are 

OFF

PURGE

  (7  liters/min), 

NORMAL

  (1  liter/min), 

and 

STANDBY

 (20 cc/min).  

 
The  Model 44-110-1 also differs in the open 
surface area of the face plate and the type of 
handle.  Window  opening  of  the  Model  44-
110-1 is 150 × 15 mm.

2. 

SENSITIVITY

 

 
 

Typical  sensitivity  of  the  Model  44-110  or 
44-110-1  is  from  30%  to  45%  4pi.  Due  to 
degradation  of  count  from  dust  in  the  open 
chamber  area,  30%  efficiency  is  stated  for 
determining  minimum  detectable  activity 
(MDA).  

 
The  MDA  (calculated  per  NUREG/  CR-
5849),  at  30%  efficiency  in  a  400  cpm 
background and a count rate instrument with 
an  approximate  22-second  response  time,  is 
approximately 503 dpm/100 cm

2

 

 

3. 

COUNTING GAS

 

 
 

Recommend  300  ft

3

  P-10  cylinder  mounted 

on a cylinder truck for mobile gas supply. 

Estimated  time  of  operation  for  one  300  ft

3

 

cylinder is approximately 40 hours. 

 

4. 

OPERATION

 

 
 
 
 
 
 
 
 
 
 
 
 
 

1. 

Connect  the  P-10  counting  gas  and 
HV/signal  cable  to  the  detector.    Place 
the  detector  on  an  uncontaminated  flat 
surface  and  turn  the  gas  on.  Check  to 
see that the switch on the detector is in 
the 

ON

  or 

PURGE

  position.  The  count 

rate will increase as the detector purges, 

until the maximum background level of 
approximately 400 cpm is reached. 

 

 

 

Purge  time  for  the  Model  44-110  is 
approximately  30  seconds  with  a  gas 
flow  rate  of  3  to  4  liters  per  minute. 
Purge  time  for  the  Model  44-110-1  is 

ATTENTION: 

Prior to use, it is necessary to remove the aluminum insert inside the gas connector. With it 
still  installed,  the  user  cannot  plug  in  the  gas  supply.  The  purpose  of  the  insert  is  to 
equalize  the  pressure  inside  the  probe  and  to  prevent  the  window  from  rupturing  during 
shipping  and  transport  in  an  aircraft.  Remove  it  by  releasing  the  locking  collar  while 
turning the probe upside down or simply pulling it out while releasing the collar. Once the 
aluminum insert is removed, the connector that is attached to the gas hose can be plugged 
in. 

 

Summary of Contents for 44-110

Page 1: ...LUDLUM MODEL 44 110 44 110 1 LARGE AREA TRITIUM DETECTORS March 2013 ...

Page 2: ...LUDLUM MODEL 44 110 44 110 1 LARGE AREA TRITIUM DETECTORS March 2013 ...

Page 3: ...anty of merchantability or fitness which extend beyond the description of the face there of If the product does not perform as warranted herein purchaser s sole remedy shall be repair or replacement at the option of Ludlum Measurements In no event will Ludlum Measurements be liable for damages lost revenue lost wages or any other incidental or consequential damages arising from the purchase use or...

Page 4: ......

Page 5: ...Model 44 110 44 110 1 Large Area Tritium Detectors TABLE OF CONTENTS 1 GENERAL 1 2 SENSITIVITY 1 3 COUNTING GAS 1 4 OPERATION 1 5 MAINTENANCE 2 6 DRAWINGS DIAGRAMS 3 ...

Page 6: ...iciency in a 400 cpm background and a count rate instrument with an approximate 22 second response time is approximately 503 dpm 100 cm2 3 COUNTING GAS Recommend 300 ft3 P 10 cylinder mounted on a cylinder truck for mobile gas supply Estimated time of operation for one 300 ft3 cylinder is approximately 40 hours 4 OPERATION 1 Connect the P 10 counting gas and HV signal cable to the detector Place t...

Page 7: ...ved and very small surfaces such as drain pipes table legs etc cannot be effectively assessed using either of these detectors NOTE In order to prevent residue forming on the chamber and anode wires the detector should not be used on very dirty or dusty surfaces Extreme humidity in the air in excess of 85 and moisture on surfaces can cause the detector to behave erratically Therefore it is not reco...

Page 8: ...NGS AND DIAGRAMS M 44 110 TRITIUM FRISKER Drawing 342 170A M 44 110 ASSY Drawing 342 170B 170D M 44 110 1 VALVE ASSEMBLY Drawing 342 171 M 44 110 1 PLUMBING ASSEMBLY Drawing 342 171A M 44 110 1 VALVE CALIBRATION Drawing 342 171B FACE PLATE M 44 110 1 Drawing 342 172 ...

Page 9: ......

Page 10: ......

Page 11: ......

Page 12: ......

Page 13: ......

Page 14: ......

Page 15: ......

Page 16: ......

Reviews: