Radon Mode Considerations
The Model 334A may be used as a radon progeny monitor, measuring and alarming on PAEC
(potential alpha energy concentration) levels of milli-Working Levels (mWL), and dose estimates
in mWL-h. Alternatively, if configured for SI units, the instrument will display radon
concentration in µJ/m³ and dose estimates in µJ-h/m³.
The PAEC reading is treated as the primary isotope-of-interest and the chronic PAEC dose is
shown as the default reading on the instrument (i.e. PAEC/C).
The PAEC is based on the individual radon decay product concentrations as measured off the
filter. The beta-decay product Pb-214 is estimated from the concentrations of Po-218 and Po-
214 before its contribution to the PAEC can be added. The thoron contributions from Po-212
and Bi-212 are also included in the PAEC result and, when Bi-212 is present, the 6.0 MeV peak
counts are corrected for known contributions from Bi-212 before the ultimate energy yield of
13.69 MeV of Po-218 are summed.
While its contribution is not included in the PAEC result, the Model 334A does monitor for and
report the concentration and dose estimate from Po-210.