4.4. SETUP MODE OPERATION
39
sired option.
- Normal:
All parameters are protected from change through the device menu. This is the
default mode when the Setup Protect dipswitch is configured to the ON position.
- Bypass Alarms:
All parameters except for Rate, Count, and Dose Alarms are protected from
change through the device menu.
- Bypass Alarms Count:
All parameters except for Rate Alarms, Count Alarms, Dose Alarms,
and the User Defined Scaler Count Time are protected from change through the device
menu.
- Bypass All:
All parameters may be changed from the Setup Menu, similar to leaving the
Setup Protect DIP Switch set to the OFF position.
J
With the DIP Switch 2 in the ON position, setup mode may be entered through
the front panel, and any parameter which has not been bypassed may be
viewed, but changes cannot be made.
Only bypassed parameters may be
changed.
DISPLAY BACKLIGHT ‘Continuous On’:
The Model 3014 display backlight can be set to
remain on continuously during operation. Follow the steps above for
SETUP PROTECT
, but
use DIP Switch 1 for display backlight selection. Setting DIP Switch 1 to the ON (forward)
position will configure the display backlight to remain on during operation. Set DIP Switch
1 to the OFF (back) position, and the display will be backlit only when light levels are low.
J
Setting the display backlight for continuous-on operation can result in reduced
battery life. The backlight can be configured to always be off as well through the
device menu or Lumic 2.0 Calibration Software.
4.4 Setup Mode Operation
Once the Model 3014 is in Setup mode, setup page selection will be displayed on the LCD.
Short press the MODE or ON/AUD button to choose the Setup Page you are interested in.
Once the desired Setup Page is shown, short press the UNIT button to cycle down to the first
parameter of that Setup Page. Further short presses on the MODE button will cycle down
through the parameters for that page. Short press the ON/AUD button to cycle up through
the parameters of that page. Extra-long-pressing the UNIT button will go back up to the
Setup Page. Extra-long pressing again will exit the Setup Mode.
March 2021