Model 3030P
Technical Manual
Section 4
Ludlum Measurements, Inc.
Page 16
February 2021
LOW BATT
: A red LED indicator illuminates whenever the instrument is
operating on batteries and the battery voltage is low.
CPM
: A green LED indicator illuminates whenever CPM mode is
established (during setup).
DPM
: A green LED indicator illuminates whenever DPM mode is
established (during setup).
αAL
: A red LED indicator illuminates whenever the alpha count exceeds the
alarm level established during setup.
βAL
: A red LED indicator illuminates whenever the beta count exceeds the
alarm level established during setup.
Back Panel
ON/OFF
: a rocker switch used to apply power to the instrument when in the
ON
position
USB
:
a type-B connector for connection to a computer
Power Input Jack
: 5.0 Vdc, 5W
2 Battery Tubes:
Each tube holds 2 “D” cell batteries.
Internal Controls
The following dipswitches and jumpers are located on the
amplifier/processor board and are only accessible when the cover is
removed:
Dipswitch 1
: When in the
OFF
position, the Model 3030P is locked into
scaler mode only. The default position is
ON
.
Dipswitch 2
: When in the
OFF
position, the calibration date is enabled. The
default position is
ON
.
Summary of Contents for 251450
Page 2: ......
Page 5: ......
Page 58: ...Model3030P TechnicalManual Section7 Ludlum Measurements Inc Page 50 February 2021 ...
Page 59: ...Model3030P TechnicalManual Section7 Ludlum Measurements Inc Page 51 February 2021 ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......