32
CHAPTER 4. SETUP MODE
4.2 Setup Parameters
Setting
Setup Parameter
Model
133-6
Model
44-2
Alpha
or Beta
Model
42-
31H
Model
44-132
P1-1
Calibration Constant
Mantissa
122
100
001
600
940
P1-2
Calibration Constant
Exponent
04
08
01
04
8
P1-3
Dead Time Correction 1
36
10
5
5
5
P1-4
Dead Time Correction 2
Mantissa
360
TBD
P1-5
Dead Time Correction 2
Exponent
-12
TBD
P1-6
Efficiency
100
15
Alpha:
20
Beta:
15
15
15
P1-7
High Voltage Set Point
550
800
800
1200
800
P1-8
Discriminator Threshold
35
10
35
2
10
P1-9
Detector Current
Overload Threshold
0 (disabled)
P1-10
Loss of Count Alarm Time
999
60
Alpha:
600
Beta:
60
0 (dis-
abled)
60
P1-11
Detector Enable
En
P2-1
Unit 1 - Units and
Minimum Display
00.0
mR/h
000
µ
R/h
000
cpm
0.00
µ
Sv/h
00.0
µ
R/h
P2-2
Unit 1 - Units Maximum
Display
999
R/h
50.0
mR/h
999
kcpm
100
mSv/h
2.50
mR/h
P2-3
Unit 1 - RATE/ MAX Mode
Alarm 1
20.0
mR/h
2.00
mR/h
2.00
kcpm
20.0
µ
Sv/h
200
µ
R/h
P2-4
Unit 1 - RATE/ MAX Mode
Alarm 2
50.0
mR/h
5.00
mR/h
5.00
kcpm
50.0
µ
Sv/h
500
µ
R/h
P2-5
Unit 1 - Count Units and
Minimum Display
00.0
mR
0.00
mR
000
counts
0.00
µ
Sv
00.0
µ
R
P2-6
Unit 1 - Count Units
Alarm 1
000 R
(dis-
abled)
000 R
(dis-
abled)
000
counts
(dis-
abled)
000 Sv
(dis-
abled)
000 R
(dis-
abled)
P2-7
Unit 1 - Count Units
Alarm 2
000 R
(dis-
abled)
000 R
(dis-
abled)
000
counts
(dis-
abled)
000 Sv
(dis-
abled)
000 R
(dis-
abled)
Ludlum Measurements, Inc.
Summary of Contents for 25009185
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 12: ......
Page 14: ......
Page 32: ......
Page 35: ...29 Weight 0 7 kg 1 6 lb excluding detectors and clips December 2021 ...
Page 36: ......
Page 56: ......
Page 62: ......
Page 72: ......
Page 74: ......
Page 78: ......