Ludlum Measurements 239-1F Manual Download Page 5

 

Model 239-1F Floor Monitor 

 December 

2010 

 

1. GENERAL

 

 
  The Model 239-1F Floor Monitor is a gas 
proportional floor monitor detector mounted on 
a roll-around cart.  The instrument features a 
flow system, quick-connects, a gas bottle 
mount, and a means to adjust the height of the 
detector from the floor for optimum 
performance.  A nuclear counting gas bottle, 
gas regulator, and a survey instrument are 
required for the operation of the floor monitor. 

 The 

survey 

instrument should be capable 

of supplying up to 2500 volts to the detector.
  The floor monitor utilizes quick-connect 
fittings on the supply side of the input flow 
meter and the external outlet line of the output 
flow meter for flow operation.  An input line to 
the flow meter with quick-connect is attached 
from the portable gas bottle mounted on the 
frame of the cart.   

 
 

2. SPECIFICATIONS

 

 

 

 DETECTOR 

SIZE

: 2 x 16 x 46.5 

cm (0.8 x 6.3 x 18.3 inch) (H x W x L)  
 

 DETECTOR 

ACTIVE 

AREA

: 584 

cm

2

 

 

 ADJUSTABLE 

HEIGHT

: Detector 

adjusts from 0 inches to 1

½ 

inches from floor 

 

 WINDOW 

MATERIAL

: (a) 0.4 or 

0.8 mg/cm

2

 (one or two layers metalized 

mylar) alpha, beta/gamma; (b) 3.9 mg/cm

2

 

(one layer metalized mylar, one layer 3.5 
mg/cm

2

 mylar) beta/gamma; (c) 7.9 mg/cm

2

 

(one layer metalized mylar, one layer 7.5 
mg/cm

2

 mylar) gamma 

 

 EFFICIENCY  (4

π

):

 17%-

239

Pu; 

25%-

90

Sr/

90

Y; gamma <1%.  Efficiencies are 

expressed in 4pi geometry and calculated with 
a probe using the 0.8 mg/cm

Mylar window 

at a fixed height of 3/16 inches distance from 
the floor

   

 

 COUNTER

: Any instrument suitable 

for use with proportional detectors and 
capable of 2500V operation.  

 

 CONNECTOR

: Series "C" type 

 

 

 QUICK-CONNECTS

: Swagelock 

(brand) 1/8 mpt to 1/4 O. D. tubing.  
 

 GAS 

CONSUMPTION

approximately 35 cc/min. 

 

 FLOW 

METERS

: Both IN and OUT  

0-100 cc/min. 
 

 FLOW RATE

: 35 cc/min after 

purging 
 

 MAXIMUM GAS CYLINDER 

SIZE

:  Size is 9 inch diameter x 26 inches; 

0.60 cu. ft. capacity 
 

 WIRE

:  0.001 inches diameter 

platinum (8 % Tungsten alloy) 
 

 STANDOFFS

:  Concord 1102-23 -

0019 
      

   

CART: 

 

 

  Handle Height: 3.5 ft. 

 

 

 Length: 66.04cm (26 inch), 

including wheels, but excluding 
handle. 

 

 

  Width: 46.36 cm (18.25 inch) 

 

Wheel Size: Rear 8 inches, front 3 
inches swivel. 

 

Finish: Computer Beige powder 
coat paint. 

 

 

WEIGHT:

 20 lbs. without gas cylinder 

and counting instrument. 

 

 

3

 

See instrument manual for specifications and instructions for the counting instrument. 

 

 

 

page 1  

Summary of Contents for 239-1F

Page 1: ...LUDLUM MODEL 239 1F FLOOR MONITOR Revised December 2010 LUDLUM MEASUREMENTS INC 501 OAK ST P O BOX 810 SWEETWATER TX 79556 325 235 5494 FAX 325 235 4672 ...

Page 2: ...anty of merchantability or fitness which extend beyond the description of the face there of If the product does not perform as warranted herein purchaser s sole remedy shall be repair or replacement at the option of Ludlum Measurements In no event will Ludlum Measurements be liable for damages lost revenue lost wages or any other incidental or consequential damages arising from the purchase use or...

Page 3: ...Model 239 1F Floor Monitor December 2010 TABLE OF CONTENTS 1 GENERAL 1 2 SPECIFICATIONS 1 3 GAS BOTTLE INSTALLATION 2 4 OPERATING PROCEDURES 2 5 MAINTENANCE 3 6 REPLACEMENT PARTS LIST 4 ...

Page 4: ...Model 239 1F Floor Monitor December 2010 Model 239 1F Floor Monitor shown with Ludlum Model 2221 ...

Page 5: ... cm2 one or two layers metalized mylar alpha beta gamma b 3 9 mg cm2 one layer metalized mylar one layer 3 5 mg cm2 mylar beta gamma c 7 9 mg cm2 one layer metalized mylar one layer 7 5 mg cm2 mylar gamma EFFICIENCY 4π 17 239 Pu 25 90 Sr 90 Y gamma 1 Efficiencies are expressed in 4pi geometry and calculated with a probe using the 0 8 mg cm2 Mylar window at a fixed height of 3 16 inches distance fr...

Page 6: ...he plastic nuts and screw the nut onto the fitting Finger tight is sufficient 6 Proceed to the operating procedures section of this manual 4 OPERATING PROCEDURES 3CAUTION If the output has a quick connect fitting insert a male quick connect prior to applying the gas supply o Turn the flow control valve off 3CAUTION Do not overtighten o Turn valve on gas supply bottle to the OPEN position The high ...

Page 7: ...ow counting gas to escape causing degradation of detector efficiency If the window needs replacing in the detector the detector should be checked both for gas leaks and HV shorts before replacing detector on the cart Indication of a broken wire is low or erratic HV reading Foreign particles wire tails on the standoff posts or solder peaks can cause excessive counts Loose wires can cause reduced co...

Page 8: ...EA Main Box Screened 9175 014 1 EA 5 Ft C Cable 8303 022 10 EA PLT2M Cable Tie 03 5410 4 EA Stem Fit B OC4 D 400 13 7912 2 EA M Pipe Fit B QC4 B 2PM 13 7920 2 EA Grommet 2170 18 8774 1 EA Flowmeter RMA 150 SSV 21 9547 1 EA Flowmeter RMA 150 21 9548 15 FT Vinyl Tubing 59002 22 9514 12 FT Swaglok B 400 1 2 22 9522 10 EA Swaglok Insert B 405 2 22 9639 28 IN Edge Trim 8451A53 22 9828 REGULATOR 2310017...

Reviews: