Touchclave Ecotech
26
Section 3- Control System Basic Screens
Note: All screens shown for illustrative purposes only
3.1 Power up/ Idle Screen
Upon initial power-up of the machine, the following screen will be displayed on the Touchscreen.
Press anywhere on the Touchscreen display to move onto home.
Note: Version number depends on software version and so may vary from image shown.
This screen will also be displayed if the touchscreen has not been touched for a period of time and
acts as a screensaver.
3.2 Process Screen
When in standby, the autoclave is out of cycle and with the control system awaiting commands from
the user to run a cycle.
Process screen displays chamber conditions- Absolute Chamber pressure, Load PT100 temperature
probe temperature and Vent PT100 temperature probe temperature. A button is displayed allowing
user to either lock or unlock the chamber door- depending on the state the door is currently in.
Buttons are also available on this screen to view any active alarms, enter the menu and to start a
cycle (note- user will have to input authorised password to start cycle, enter menu or acknowledge
any alarms).
Current Time is also displayed on process screen.
Door can be locked or unlocked without password entry.
V1.0.0